H Cloete1, W Kleinhans, E P G Mansvelt.
Abstract
Ferritin assays were performed using both serum and plasma samples on the Bayer Technicon Immuno 1 Analyser and their ferritin test kit as only data for serum samples are available. After evaluation of the test kit and analyser the results were correlated using the Statistica for Windows programme. It was established that the plasma samples gave consistently lower values than the serum samples. Using the obtained conversion factor of 1.191, plasma as well as serum samples may be used for ferritin determinations.Mesh:
Substances:
Year: 2002 PMID: 12358888 DOI: 10.1046/j.1365-2257.2002.00469.x
Source DB: PubMed Journal: Clin Lab Haematol ISSN: 0141-9854