The nucleus contour electrode array: an electrophysiological study.

Enrico Pasanisi1, Vincenzo Vincenti, Andrea Bacciu, Maurizio Guida, Salvatore Bacciu.   

Abstract

OBJECTIVE/HYPOTHESIS: Positioning intracochlear electrodes adjacent to the modiolus would lead to a reduction in threshold levels and latencies, as measured with electrically evoked auditory brainstems responses. Recently, Cochlear Corporation introduced a new perimodiolar array (Contour electrode) characterized by a platinum stylet inserted into the array's silicone carrier that keeps the electrode in a straightened configuration before insertion. After completion of the insertion, the withdrawal of the stylet allows the array to coil within the cochlea positioning electrodes closer to the modiolus. The purpose of the study was to analyze the stimulation properties of the Contour electrode by evaluating the effects of the withdrawal of the stylet on electrical auditory brainstem response (EABR) characteristics.
METHODS: Electrical auditory brainstem response registrations were performed intraoperatively on three electrodes (1 apical, 1 medial, 1 basal) before and after the removal of the stylet in 10 patients implanted with the Nucleus 24 Contour device. A within-patient comparison of EABR characteristics was performed. The Student test was used to assess differences in EABR measures obtained before and after the stylet removal.
RESULTS: Mean EABR thresholds were significantly lower after the removal of the stylet than with the stylet in situ for all three electrodes tested. After the removal of the stylet, latencies of waves III and V significantly decreased, whereas amplitudes significantly increased.
CONCLUSION: Reduction in EABR threshold levels and latencies as well as an increase in waves III and V amplitudes after the stylet removal all demonstrate the effectiveness of the Contour system in positioning intracochlear electrodes closer to auditory neurons.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12352681     DOI: 10.1097/00005537-200209000-00023

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  2 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

2.  Modiolus-hugging intracochlear electrode array with shape memory alloy.

Authors:  Kyou Sik Min; Sang Beom Jun; Yoon Seob Lim; Se-Ik Park; Sung June Kim
Journal:  Comput Math Methods Med       Date:  2013-05-09       Impact factor: 2.238

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.