| Literature DB >> 12270234 |
K J M Surry1, W L Smith, L J Campbell, G R Mills, D B Downey, A Fenster.
Abstract
We have designed a prototype three-dimensional ultrasound guidance (3D USB) apparatus to improve the breast biopsy procedure. Features from stereotactic mammography and free-hand US-guided biopsy have been combined with 3D US imaging. This breast biopsy apparatus accurately guides a needle into position for the sampling of target tissue. We have evaluated this apparatus in three stages. First, by testing the placement accuracy of a needle in a tissue mimicking phantom. Second, with tissue mimicking phantoms that had embedded lesions for biopsy. Finally, by comparison to free-hand US-guided biopsy, using chicken breast phantoms. The first two stages of evaluation quantified the mechanical biases in the 3D USB apparatus. Compensating for these, a 96% success rate in targeting 3.2 mm "lesions" in chicken breast phantoms was achieved when using the 3D USB apparatus. The expert radiologists performing biopsies with free-hand US guidance achieved a 94.5% success rate. This has proven an equivalence between our apparatus, operated by non-experts, and free-hand biopsy performed by expert radiologists, for 3.2 mm lesions in vitro, with a 95% confidence.Entities:
Mesh:
Year: 2002 PMID: 12270234 DOI: 10.1016/s1361-8415(02)00087-7
Source DB: PubMed Journal: Med Image Anal ISSN: 1361-8415 Impact factor: 8.545