Literature DB >> 12270231

A CAD system for the 3D location of lesions in mammograms.

Yasuyo Kita1, Eriko Tohno, Ralph Highnam, Michael Brady.   

Abstract

A CAD system for estimating the 3D (three-dimensional) positions of lesions found in two mammographic views is described. The system is an extension of our previous method [Comput. Vis. Image Understand. 83 (2001) 38] which finds corresponding 2D positions in different mammographic views. The method calculates curved epipolar lines by developing a simulation of breast deformation into stereo camera geometry. Using such curved epipolar lines, not only can we determine point correspondences, but can estimate the 3D location of a lesion. In this paper, we first explain the underlying principles and system organisation. The correctness of the 3D positions calculated by the system is examined using a set of breast lesions, which appear both in mammograms and in MRI data. The experimental results demonstrate the clinical promise of the CAD system.

Entities:  

Mesh:

Year:  2002        PMID: 12270231     DOI: 10.1016/s1361-8415(02)00084-1

Source DB:  PubMed          Journal:  Med Image Anal        ISSN: 1361-8415            Impact factor:   8.545


  1 in total

1.  Using x-ray mammograms to assist in microwave breast image interpretation.

Authors:  Charlotte Curtis; Richard Frayne; Elise Fear
Journal:  Int J Biomed Imaging       Date:  2012-03-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.