Literature DB >> 12243442

Electrical failure during cardiopulmonary bypass: an evaluation of incidence, causes, management and guidelines for preventative measures.

M Hargrove1, B C Ramish, A O'Donnell, T Aherne.   

Abstract

The incidence of electrical failure during cardiopulmonary bypass (CPB) has been reported to occur in approximately 1 per 1000 cases. While the resultant morbidity and mortality is low, electrical failure is a life-threatening scenario. We report three major electrical failures during CPB in a patient population of 3500 over a 15-year period. These cases involved mains failure and generator shut down, mains failure and generator power surge, and failure of the uninterruptable power supply (UPS), which caused protected sockets to shut down. Protocols for preventative maintenance, necessary equipment, battery backup and guidelines for the successful management of such accidents during CPB are discussed.

Entities:  

Mesh:

Year:  2002        PMID: 12243442     DOI: 10.1191/0267659102pf592oa

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  6 in total

Review 1.  Crisis in the operating room: fires, explosions and electrical accidents.

Authors:  Keiko Nishiyama; Makiko Komori; Mitsuharu Kodaka; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2010-08-14       Impact factor: 1.731

2.  Variations in battery life of a heart-lung machine using different pump speeds, pressure loads, boot material, centrifugal pump head, multiple pump usage, and battery age.

Authors:  Cornelius Marshall; Martin Hargrove; Aonghus O'Donnell; Thomas Aherne
Journal:  J Extra Corpor Technol       Date:  2005-09

3.  Blackout of 2003: public health effects and emergency response.

Authors:  Mark E Beatty; Scot Phelps; Mpha Chris Rohner; Mupa Isaac Weisfuse
Journal:  Public Health Rep       Date:  2006 Jan-Feb       Impact factor: 2.792

4.  Possible fire hazard caused by mismatching electrical chargers with the incorrect device within the operating room.

Authors:  Martin Hargrove; Thomas Aherne
Journal:  J Extra Corpor Technol       Date:  2007-09

5.  Quantitative evaluation of hand cranking a roller pump in a crisis management drill.

Authors:  Yasuko Tomizawa; Asako Tokumine; Shinji Ninomiya; Naoki Momose; Toru Matayoshi
Journal:  J Artif Organs       Date:  2008-10-05       Impact factor: 1.731

6.  Electrical failure during cardiopulmonary bypass: a critical moment.

Authors:  Ahmet Baris Durukan; Hasan Alper Gurbuz; Gokhan Ozcelik; Cem Yorgancioglu
Journal:  Kardiochir Torakochirurgia Pol       Date:  2016-06-30
  6 in total