The electronic portal imaging system Siemens Beamview Plus versus the conventional verification films CEA-TVS and DuPont COL-7. A critical appraisal of visual image quality.

Bernd Gagel1, Oliver Schramm, Wolfgang Harms, Andrew Mulhern, Frederik Wenz, Michael van Kampen, Michael Wannenmacher, Michael J Eble.   

Abstract

AIM: The aim of this study was the validation of the visual image quality of electronic portal imaging devices (EPID) and conventional verification films from the point of view of the end-viewers of portal films, the radiotherapists.
MATERIAL AND METHODS: The verification image was represented in two different forms, viz. an electronic portal image employing Siemens Beamview Plus (on a computer monitor) and two different portal films using the conventional verification films CEA-TVS and DuPont CQL-7 (on a negatoscope). A total of 270 image sets (simulation film and portal image) were evaluated by each radiotherapist, evaluation extending to 90 sets of each type of verification film. Each set was evaluated by three specialists in radiotherapy examining subjective visual image quality whereby the following aspects served as evaluation criteria: contrast, artifacts, determination of actual radiation field edge position, anatomical structures and main structural feature for the determination of treatment field position. In addition, the anatomical structures employed for visual feature correlation between reference and portal films were classified according to their importance.
RESULTS: In general the electronic portal image was rated significantly "visible" or better. Only the evaluation of artifacts showed an appreciable disadvantage for electronic portal imaging caused by physical artifacts due to radiographic technique and data processing aspects peculiar to the Siemens Beamview Plus 1.1. and also caused by different image processing tools reducing physical artifacts and enhancing the visibility of anatomical structures and likewise of anatomical artifacts (e.g. intestinal gas). By calculating the Spearman correlation coefficient to detect a possible relationship between the different criteria of subjective visual image quality, the research demonstrated that artifacts when limited to a tolerable proportion had no significant impact on the other criteria.
CONCLUSIONS: As data of EPIDS are digital, images can be postprocessed and enhanced in a wide variety of ways. Using this tool the electronic portal imaging device provides images that, in terms of visual image quality, are at least comparable to the two evaluated types of radiographic films and also have the added advantage that such images are stored and can be transferred electronically being presupposition for digital patient documentation.

Entities:  

Mesh:

Year:  2002        PMID: 12240551     DOI: 10.1007/s00066-002-0964-1

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  5 in total

1.  Abstracts of the 1st International Conference on the Clinical Use of Tomotherapy. October 17-18, 2008. Munich, Germany.

Authors: 
Journal:  Strahlenther Onkol       Date:  2009-01       Impact factor: 3.621

2.  Third party EPID with IGRT capability retrofitted onto an existing medical linear accelerator.

Authors:  D O Odero; D S Shimm
Journal:  Biomed Imaging Interv J       Date:  2009-07-01

3.  Evaluation of three presets for four-dimensional cone beam CT in lung radiotherapy verification by visual grading analysis.

Authors:  Sally A Kember; Vibeke N Hansen; Martin F Fast; Simeon Nill; Fiona McDonald; Merina Ahmed; Karen Thomas; Helen A McNair
Journal:  Br J Radiol       Date:  2016-04-25       Impact factor: 3.039

4.  A quality assurance phantom for electronic portal imaging devices.

Authors:  Indra J Das; Minsong Cao; Chee-Wai Cheng; Vladimir Misic; Klaus Scheuring; Edmund Schüle; Peter A S Johnstone
Journal:  J Appl Clin Med Phys       Date:  2011-02-02       Impact factor: 2.102

5.  Potentials of on-line repositioning based on implanted fiducial markers and electronic portal imaging in prostate cancer radiotherapy.

Authors:  Reinhold Graf; Peter Wust; Volker Budach; Dirk Boehmer
Journal:  Radiat Oncol       Date:  2009-04-27       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.