Topography-based screening for previous laser in situ keratomileusis to correct myopia.

François Destrempes1, Isabelle Brunette, Jean Meunier, Mylène Beyrouthy, Pierre Demers, Sameh Fanous, Guy Doyon.   

Abstract

PURPOSE: To demonstrate the feasibility of developing a screening tool based on corneal topography to detect previous myopic laser in situ keratomileusis (LASIK).
SETTING: Clinical data from a private clinic analyzed in a university setting.
METHODS: Two hundred thirty-three topographies were randomly selected so 1 topography per patient was used: 150 from unoperated corneas and 83 from corneas that had LASIK to correct myopia. The mean surgical correction was -4.40 diopters (D) +/- 2.53 (SD) (range -11.00 to -0.38 D). All topographies were performed using an Orbscan II unit (Bausch & Lomb Surgical). The LASIK procedures were performed using a Technolas 217C excimer laser and a Hansatome microkeratome (Bausch & Lomb Surgical). The algorithms used the mean value of the directional derivative (DT) of the anterior tangential curvature of the cornea in the 2.2 mm radius central disk and the mean value of the anterior elevation (E) with respect to the best-fit sphere in the 0.5 mm radius central disk. Topographies in the testing set (n = 119) were classified as operated if E < 0 (E algorithm) or DT > 0 (DT algorithm) or as unoperated.
RESULTS: The E algorithm yielded 0% false positives and 16.7% false negatives and the DT algorithm, 6.5% and 7.1%, respectively. For myopia greater than -1.12 D, the DT algorithm provided a 0% false negative rate. The performance of E and DT algorithms, used in combination, was superior to clinical assessment.
CONCLUSION: Criteria based on Orbscan II corneal topography are proposed for the detection of previous myopic LASIK performed with a Technolas 217C excimer laser.

Entities:  

Mesh:

Year:  2002        PMID: 12231326     DOI: 10.1016/s0886-3350(02)01267-1

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Anterior elevation maps as the screening test for the ablation power of previous myopic refractive surgery.

Authors:  Soo Yong Jeong; Hee-Seung Chin; Jung Hyub Oh
Journal:  Korean J Ophthalmol       Date:  2006-03