Thomas B Martins1, Brian M Pasi, Jerry W Pickering, Troy D Jaskowski, Christine M Litwin, Harry R Hill.
Abstract
We used a multiplexedfluorescent microsphere immunoassay to develop a sandwich capture assay to assess simultaneously the production of thymus helper (TH) 1- and TH2-type cytokines in tissue culture supernatant obtained from stimulated peripheral blood mononuclear cells. The assay then was used to assess the cytokine production of patients with hyperimmunoglobulinemia E syndrome and in cord blood from neonates. The multiplexed assay has a reportable range of less than 10 to 50,000 pg/mL. For linearity and recovery studies, R2 values for the 6 cytokines ranged from 0.988 to 0.999 for samples spiked with known concentrations of recombinant cytokine standards and for patient samples. The assay showed good specificity, with little cross-reactivity between cytokines. Results from supernatants of Staphylococcus aureus-stimulated peripheral blood mononuclear cells obtainedfrom 6 patients with hyperimmunoglobulinemia E syndrome showed significantly less interferon (IFN)-gamma production than cells from healthy control subjects. Cord blood cells from neonates produced significantly less interleukin 12 and IFN-gamma than cells from adults in group B streptococci-stimulated mononuclear cells. The fluorescent multiplexed microsphere immunoassay can be used to quantitate multiple cytokines from 1 sample and should be useful for further understanding of the cytokine role in disease.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12219776 DOI: 10.1309/N0T6-C56B-GXB2-NVFB
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493