The accuracy of the i-STAT portable analyser for measuring blood gases and pH in whole-blood samples from dogs.

P Verwaerde1, C Malet, M Lagente, F de la Farge, J P Braun.   

Abstract

To assess the suitability of the i-STAT portable analyser for use by non-laboratory personnel, we measured blood gases and pH in venous blood samples from 100 dogs. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using two different autocalibrated benchtop analysers. Overall accuracy of the portable analyser proved excellent for pH, pO(2), and pCO(2) (r=0.978, 0.968 and 0.997, respectively), with Deming's regression slopes close to 1.00 (0.96, 0.97 and 1.08 for pH, pO(2), and pCO(2), respectively) and intercepts close to zero (0.28, 0.47 kPa and 0.46 kPa for pH, pO(2), and pCO(2), respectively). The accuracy of the i-STAT was also satisfactory for calculated parameters: bicarbonates, total CO(2), base excess and oxygen saturation. Our findings show this portable analyser to be a valid substitute for expensive benchtop analysers in situations requiring mobility, or when small numbers of tests are to be performed by users not specialized in laboratory techniques.

Entities:  

Mesh:

Year:  2002        PMID: 12208109     DOI: 10.1016/s0034-5288(02)00065-6

Source DB:  PubMed          Journal:  Res Vet Sci        ISSN: 0034-5288            Impact factor:   2.534


  5 in total

1.  Influence of grain processing on acid-base balance in feedlot steers.

Authors:  C Castillo; J Hernández; J Méndez; J Llena; V Pereira; M López-Alonso; J L Benedito
Journal:  Vet Res Commun       Date:  2006-10       Impact factor: 2.459

2.  Measurement of blood-oxygen saturation using a photoacoustic technique in the rabbit hypoxemia model.

Authors:  Kiguna Sei; Masanori Fujita; Takeshi Hirasawa; Shinpei Okawa; Toshihiro Kushibiki; Hidenori Sasa; Kenichi Furuya; Miya Ishihara
Journal:  J Clin Monit Comput       Date:  2018-06-06       Impact factor: 2.502

3.  Evaluation of the i-STAT (portable clinical analyser) for measuring haematological parameters in Atlantic cod (Gadus morhua) at different CO2 and temperature conditions.

Authors:  Radoslav S Borissov; Sølvi Espeland; Martin H Iversen
Journal:  Fish Physiol Biochem       Date:  2019-10-03       Impact factor: 2.794

4.  Intraoperative measurement of bowel oxygen saturation using a multispectral imaging laparoscope.

Authors:  Neil T Clancy; Shobhit Arya; Danail Stoyanov; Mohan Singh; George B Hanna; Daniel S Elson
Journal:  Biomed Opt Express       Date:  2015-09-30       Impact factor: 3.732

5.  Evaluation of the i-STAT Alinity v in a veterinary clinical setting.

Authors:  Jasper E Burke; Thuy Hien T Nguyen; Tanya Davis; Amie Koenig; Selena L Lane; Jennifer Good; Benjamin M Brainard
Journal:  J Vet Diagn Invest       Date:  2021-05-28       Impact factor: 1.569

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.