Use of a fast EM algorithm for 3D image reconstruction with the YAP-PET tomograph.

A Motta1, C Damiani, A Del Guerra, G Di Domenico, G Zavattini.   

Abstract

OBJECTIVE: We would like to improve the image reconstructions for both signal-to-noise ratio (SNR) and spatial resolution characteristics for the small animal positron emission tomograph YAP-PET, built at the Department of Physics of Ferrara University. The three-dimensional (3D) filtered backprojection (FBP) algorithm, usually used for image reconstruction, has a limited angle restriction due to the tomograph geometry, which causes a serious loss in sensitivity.
METHODS: We implemented a 3D iterative reconstruction program using the symmetry and sparse properties of the 'probability matrix', which correlates the emission from each voxel to the detector within a coincidence tube. A fraction only of matrix elements are calculated before the reconstruction and stored on disk: this allows us to avoid on-line computation. A depth dependent function differentiates the voxels in a coincidence tube. Three experimental phantoms with no background were reconstructed by using the program, in comparison with traditionally used FBP.
RESULTS: The adopted method allowed us to reduce the computation time significantly. Furthermore, the simple depth dependent function improved the spatial resolution. With 64 x 64 x 20 voxels of 0.625 x 0.625 x 2.0 mm(3) in the field of view, the computation time was less than 4 min per iteration on a Sparc Ultra 450 Workstation, and less than 6 min per iteration on a Mac-PPC G3 300 MHz: the spatial resolution measured with a 0.8 mm diameter 18F-FDG filled capillary reconstructed in this way was 2.0 mm FWHM. By decreasing the voxel size to 0.3125 x 0.3125 x 2.0 mm(3) per voxel the transaxial FWHM was 1.7 mm with a computation time of 15 min per iteration on a Sparc Ultra 450. By using all the acquired data, the SNR improves from 1.3 to 6.0 in the worst measured case, a pair of 0.8mm diameter 18F-FDG filled capillaries, which are 2.5 mm apart each other.
CONCLUSION: The adoption of iterative reconstruction allowed us to overcome the loss in sensitivity of previously used FBP: this improved the SNR. The studies of symmetry and sparse properties avoided a severe increase of the reconstruction time and of storing space on disk. This fast EM Algorithm is now routinely used for the image reconstruction with the YAP-PET tomograph.

Entities:  

Mesh:

Year:  2002        PMID: 12204233     DOI: 10.1016/s0895-6111(02)00034-4

Source DB:  PubMed          Journal:  Comput Med Imaging Graph        ISSN: 0895-6111            Impact factor:   4.790


  4 in total

Review 1.  From PET detectors to PET scanners.

Authors:  John L Humm; Anatoly Rosenfeld; Alberto Del Guerra
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-10-02       Impact factor: 9.236

2.  A novel AMPK activator reduces glucose uptake and inhibits tumor progression in a mouse xenograft model of colorectal cancer.

Authors:  Silvia Valtorta; Gabriella Nicolini; Farida Tripodi; Cristina Meregalli; Guido Cavaletti; Federica Avezza; Luca Crippa; Gloria Bertoli; Francesca Sanvito; Paola Fusi; Roberto Pagliarin; Fulvia Orsini; Rosa Maria Moresco; Paola Coccetti
Journal:  Invest New Drugs       Date:  2014-08-19       Impact factor: 3.850

3.  Chronic Δ⁹-tetrahydrocannabinol exposure induces a sensitization of dopamine D₂/₃ receptors in the mesoaccumbens and nigrostriatal systems.

Authors:  Nathalie Ginovart; Benjamin B Tournier; Marcelle Moulin-Sallanon; Thierry Steimer; Vicente Ibanez; Philippe Millet
Journal:  Neuropsychopharmacology       Date:  2012-06-13       Impact factor: 7.853

4.  Characterization of preclinical models of prostate cancer using PET-based molecular imaging.

Authors:  Sara Belloli; Elena Jachetti; Rosa M Moresco; Maria Picchio; Michela Lecchi; Silvia Valtorta; Massimo Freschi; Rodrigo Hess Michelini; Matteo Bellone; Ferruccio Fazio
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03-11       Impact factor: 9.236

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.