| Literature DB >> 12201455 |
Andreas Austeng1, Sverre Holm.
Abstract
To be able to describe more precisely the behavior of a real-time 3-D ultrasound system with either a dense array or various sparse designs, experimental data from a 2-D fully connected array prototype with 50 x 50 elements have been collected. The data have been processed off line to form synthetic aperture 3-D volume images. Simulated and experimental results are compared and show good correlation. The performance of the best sparse designs, all thinned to more than 50%, offer performances comparable to a dense array.Mesh:
Year: 2002 PMID: 12201455 DOI: 10.1109/tuffc.2002.1026020
Source DB: PubMed Journal: IEEE Trans Ultrason Ferroelectr Freq Control ISSN: 0885-3010 Impact factor: 2.725