Literature DB >> 12197599

High incidence of device-related and lead-related complications in the dual-chamber implantable cardioverter defibrillator compared with the single-chamber version.

Tamana Takahashi1, Anil K Bhandari, Masato Watanuki, David S Cannom, Harumizu Sakurada, Masayasu Hiraoka.   

Abstract

Recently, the routine use of dual-chamber implantable cardioverter defibrillators (DC-ICD) has been advocated over the single-chamber version (SC-ICD), but there are few reports of the frequency of complications between the 2 types of ICDs. Between July 1997 and April 1999, 178 consecutive patients underwent implantation of either a transvenous SC-ICDs (n=104) or a DC-ICDs (n=74). Twelve (16%) of the 74 patients with a DC-ICD had a total of 16 major complications compared with 6 (6%) of the 104 patients with a SC-ICD (p=0.01). The 16 DC-ICD complications included atrial lead dislodgment (4), ventricular lead malfunction (4), and pocket infection/hematoma (3), and the 6 SC-ICD complications included ventricular lead dislodgment (2) and pocket hematoma (3). Patients with a DC-ICD had less left ventricular function (29% vs 35%, p=0.035) and a higher prevalence of non-ischemic cardiomyopathies (48% vs 28%, p=0.0076). In conclusion, the DC-ICD may have a higher frequency of device- and lead-related major complications.

Entities:  

Mesh:

Year:  2002        PMID: 12197599     DOI: 10.1253/circj.66.746

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  10 in total

1.  Implantable electrophysiologic cardiac device infections: a risk factor analysis.

Authors:  D Raad; J Irani; E G Akl; S Choueiri; E Azar; J Abboud; C Afif
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2012-08-25       Impact factor: 3.267

Review 2.  Are dual-chamber implantable cardioverter-defibrillators really better than single-chamber ones? A systematic review and meta-analysis.

Authors:  Bing-Wei Chen; Qing Liu; Xu Wang; Ai-Min Dang
Journal:  J Interv Card Electrophysiol       Date:  2014-02-16       Impact factor: 1.900

3.  Delayed perforation of the right ventricular wall by a single standard-caliber implantable cardioverter-defibrillator lead detected by multidetector computed tomography.

Authors:  An Yoshimori; Atsushi Kobori; Nasu Michihiro; Yutaka Furukawa
Journal:  Korean Circ J       Date:  2011-11-29       Impact factor: 3.243

4.  Outcomes of patients with heart failure after primary prevention ICD unit generator replacement.

Authors:  Khang-Li Looi; Andrew Gavin; Lisa Cooper; Liane Dawson; Debbie Slipper; Nigel Lever
Journal:  Heart Asia       Date:  2019-02-28

Review 5.  Atrial Fibrillation in Inherited Channelopathies.

Authors:  Baha'a Al-Azaam; Dawood Darbar
Journal:  Card Electrophysiol Clin       Date:  2021-01-08

6.  Impact of body mass index on the development of pocket hematoma: A retrospective study in Chinese people.

Authors:  Jian-Ping Guo; Zhao-Liang Shan; Hong-Yang Guo; Hong-Tao Yuan; Kun Lin; Yue-Xiang Zhao; Yu-Tang Wang
Journal:  J Geriatr Cardiol       Date:  2014-09       Impact factor: 3.327

7.  Incidence, predictors and outcomes of hematoma after ICD implantation: An analysis of a nationwide database of 85,276 patients.

Authors:  Arun Raghav Mahankali Sridhar; Vivek Yarlagadda; Arun Kanmanthareddy; Sravanthi Parasa; Ryan Maybrook; Buddhadeb Dawn; Yeruva Madhu Reddy; Dhanunjaya Lakkireddy
Journal:  Indian Pacing Electrophysiol J       Date:  2016-10-21

Review 8.  ICD discrimination of SVT versus VT with 1:1 V-A conduction: A review of the literature.

Authors:  Rhanderson N Cardoso; Chris Healy; Juan Viles-Gonzalez; James O Coffey
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-12

9.  The efficacy of the LinoxSmart DX ICD lead from a single center experience.

Authors:  Ibrahim Marai; Anat Milman; Rafael Diamante; Osnat Gurevitz; David Barlev; Igor Lipchenka; Eyal Nof; Michael Glikson; Roy Beinart
Journal:  Indian Pacing Electrophysiol J       Date:  2019-12-17

10.  Unusual case of late dislodgment of a superior vena cava coil.

Authors:  Taya V Glotzer; Monali Shah; Shashank Jain; John M Zimmerman
Journal:  HeartRhythm Case Rep       Date:  2018-06-11
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.