Determination of lead in whole blood: comparison of the LeadCare blood lead testing system with Zeeman longitudinal electrothermal atomic absorption spectrometry.

Alain Pineau1, Bernard Fauconneau, Marina Rafael, Anne Viallefont, Olivier Guillard.   

Abstract

This study compares the efficiency of blood lead level analysis by graphite furnace atomic absorption spectrometry (GFAAS) and the portable LeadCare Blood Lead Testing System (LCS). Recoveries of two added lead concentrations of 22 and 42 micrograms/dL ranged from 102.4 to 105.5% for LCS and from 96.3 to 97.2% for GFAAS. Measurement of a certified sample (Certified Danish Whole Blood) at a blood lead concentration of 26.2 micrograms/dL gave within- and between-run coefficients of variation which were both approximately 8% by LCS and 2% by GFAAS. Comparison of the tested method (LCS) versus GFAAS from analysis of 76 samples of blood lead collected from workers in different industrial sectors showed imperfect overall correlation (r = 0.95). The LCS is quite suitable for screening purposes, but requires the use of non-frozen blood collected less than 24 h before. Conservative threshold values should be applied when using the LCS for initial screening in the field.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12195725     DOI: 10.1016/S0946-672X(02)80037-2

Source DB:  PubMed          Journal:  J Trace Elem Med Biol        ISSN: 0946-672X            Impact factor:   3.849


  5 in total

1.  Occupational exposure to neurotoxic substances in Asian countries - challenges and approaches.

Authors:  Monika Meyer-Baron; Eun A Kim; Iman Nuwayhid; Gaku Ichihara; Seong-Kyu Kang
Journal:  Neurotoxicology       Date:  2011-12-20       Impact factor: 4.294

2.  A Bland-Altman comparison of the Lead Care® System and inductively coupled plasma mass spectrometry for detecting low-level lead in child whole blood samples.

Authors:  Christina Sobin; Natali Parisi; Tanner Schaub; Eva de la Riva
Journal:  J Med Toxicol       Date:  2011-03

3.  Improving Reproducibility of Lab-on-a-Chip Sensor with Bismuth Working Electrode for Determining Zn in Serum by Anodic Stripping Voltammetry.

Authors:  Xing Pei; Wenjing Kang; Wei Yue; Adam Bange; William R Heineman; Ian Papautsky
Journal:  J Electrochem Soc       Date:  2014-02-01       Impact factor: 4.316

4.  Childhood correlates of blood lead levels in Mumbai and Delhi.

Authors:  Nitin B Jain; Howard Hu
Journal:  Environ Health Perspect       Date:  2006-03       Impact factor: 9.031

5.  Childhood lead exposure in the palestinian authority, Israel, and Jordan: results from the Middle Eastern regional cooperation project, 1996-2000.

Authors:  Jamal Safi; Alf Fischbein; Sameer El Haj; Ramzi Sansour; Madi Jaghabir; Mohammed Abu Hashish; Hassan Suleiman; Nimer Safi; Abed Abu-Hamda; Joyce K Witt; Efim Platkov; Steven Reingold; Amber Alayyan; Tamar Berman; Matti Bercovitch; Yogesh Choudhri; Elihu D Richter
Journal:  Environ Health Perspect       Date:  2006-06       Impact factor: 9.031

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.