Christa Cobbaert1, Cas Weykamp, Henk Baadenhuijsen, Aldy Kuypers, Jan Lindemans, Rob Jansen.   

Abstract

BACKGROUND: The Dutch project "Calibration 2000" aims at harmonization of laboratory results via calibration by development of matrix-based secondary reference materials. We considered the selection, preparation, and characterization of 34 potential reference materials (PRMs).
METHODS: Sixteen PRMs were prepared either strictly according to the NCCLS C37-A protocol or in a less stringent and more convenient way. In addition, 18 commercial, so-called human serum-based calibrators or controls were purchased and tested. Lipoprotein integrity was evaluated by examining the physicochemical characteristics of the materials. Commutability of the PRMs was assessed in 86 Dutch clinical laboratories, using a multicenter split-patient-sample between-field-method (twin-study) design. Normalized residuals of the PRMs with respect to the patient regression lines were calculated; in addition, the extra contribution of each PRM to the total measurement uncertainty (CV(Netto)) was calculated. On the basis of these results, the most native PRM was selected to investigate its potential to reduce interlaboratory variation and to improve lipid and apolipoprotein standardization.
RESULTS: In general, only the NCCLS C37-A-type materials displayed normalized residuals below the decision limit for commutability and had small CV(Netto) values ranging between 0 and 3.8%. This contrasts with the findings in regularly pooled frozen sera and lyophilized cryoprotected PRMs. In two subsequent external quality assessment surveys, the NCCLS type C37-A materials contributed to reducing the intermethod lipid and (apo)lipoprotein variation to approximately 2-4%.
CONCLUSIONS: NCCLS C37-A materials have a strong potential as secondary reference materials, not only for cholesterol but also for HDL-cholesterol, LDL-cholesterol, triglyceride, and apolipoprotein measurements.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12194930

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  Reference materials and commutability.

Authors:  Hubert W Vesper; W Gregory Miller; Gary L Myers
Journal:  Clin Biochem Rev       Date:  2007-11

2.  Reference distributions for apolipoproteins AI and B and B/AI ratios: comparison of a large cohort to the world's literature.

Authors:  Robert F Ritchie; Glenn E Palomaki; Louis M Neveux; Thomas B Ledue; Santica Marcovina; Olga Navolotskaia
Journal:  J Clin Lab Anal       Date:  2006       Impact factor: 2.352

3.  Commutability Assessment of Candidate External Quality Assessment Materials for Aminotransferase Activity Measurements Based on Different Approaches in China.

Authors:  Qichen Long; Tianqi Qi; Tianjiao Zhang; Jing Wang; Jie Zeng; Ying Yan; Meng Wang; Wei Huang; Haijian Zhao; Wenxiang Chen; Chuanbao Zhang
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

4.  Associations between smoking, components of metabolic syndrome and lipoprotein particle size.

Authors:  Sandra N Slagter; Jana V van Vliet-Ostaptchouk; Judith M Vonk; H Marike Boezen; Robin P F Dullaart; Anneke C Muller Kobold; Edith J Feskens; André P van Beek; Melanie M van der Klauw; Bruce H R Wolffenbuttel
Journal:  BMC Med       Date:  2013-09-03       Impact factor: 8.775

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.