In vivo determination of root canal length: a preliminary report using the Tri Auto ZX apex-locating handpiece.

F Grimberg1, G Banegas, L Chiacchio, O Zmener.   

Abstract

AIM: The aim of this study was to assess the clinical perfomance of a cordless handpiece with a built-in apex locator - the Tri Auto ZX - designed for root canal preparation with nickel-titanium rotary files.
METHODOLOGY: Twenty-five human maxillary incisor and canine teeth scheduled for extraction with mature apices were selected for the study. Informed written consent was obtained from each patient before treatment. After administration of local anaesthesia, the teeth were isolated and the pulp cavities accessed. The Tri Auto ZX along with a size 15 K-file was used in its electronic apex locating function based on the manufacturer's recommendations. A periapical radiograph with the file at the electronically determined constriction was taken, the file removed and the measurement registered as the electronic length (EL). To test the auto reverse function, a size 20 ProFile.04 taper NiTi rotary instrument was mounted in the handpiece. The point for the auto apical reverse function was preset on the panel at the 0.5 mm level. After the file was introduced into the canal and reached the predetermined level, the file automatically stopped and rotated in the opposite direction. A reference point was marked and this measurement was registered as the auto reverse length (ARL). All measurements were made twice by two different investigators. Teeth were then extracted and immersed in a 20% formalin solution for 48 h. After fixation, a size 15 file was inserted into the canal to measure the actual root canal length from the same reference point obtained with the Tri Auto ZX to the apical foramen, as seen in the stereo microscope. When the file tip was visible at the anatomical end of the canal it was withdrawn 0.5 mm and this measurement was registered as the actual length (AL). All measurements were expressed in mm and the measuring accuracy was set to 0.5 mm. The significance of the mean differences between EL and ARL and between EL and AL measurements at the 5% confidence level was evaluated.
RESULTS: EL measurements were coincident to ARL in all instances. EL and ARL were coincident to AL in 10 (40%) canals, in the remaining 15 canals (60%) the AL measurements were longer than EL and ARL (+0.5 mm) in 14 instances and shorter (-0.5 mm) in one case. Overall, the AL was longer than the EL or ARL, the mean difference being -0.23 mm +/- 0.32 (P < 0.05).
CONCLUSIONS: It was concluded that the Tri Auto ZX was useful and reliable. The Tri Auto ZX measurements protected against overpreparation.

Entities:  

Mesh:

Year:  2002        PMID: 12190898     DOI: 10.1046/j.1365-2591.2002.00505.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  An ex vivo comparison of working length determination by three electronic root canal length measurement devices integrated into endodontic rotary motors.

Authors:  Manal Maree Ali; Ronald Wigler; Shaul Lin; Arieh Y Kaufman
Journal:  Clin Oral Investig       Date:  2016-07-09       Impact factor: 3.573

2.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

3.  Evaluation of three instrumentation techniques at the precision of apical stop and apical sealing of obturation.

Authors:  Özgür Genç; Tayfun Alaçam; Guven Kayaoglu
Journal:  J Appl Oral Sci       Date:  2011-06-03       Impact factor: 2.698

4.  The accuracy of the auto-stop function of different endodontic devices in detecting the apical constriction.

Authors:  David W Christofzik; Andreas Bartols; Mahmoud Khaled; Birte Größner-Schreiber; Christof E Dörfer
Journal:  BMC Oral Health       Date:  2017-11-29       Impact factor: 2.757

5.  Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: a clinical study [ISRCTN71486641].

Authors:  L Smadi
Journal:  BMC Oral Health       Date:  2006-02-11       Impact factor: 2.757

6.  An in vitro stereomicroscopic comparative evaluation of a combination of apex locator and endodontic motor with an integrated endodontic motor.

Authors:  Ch Swarupa; Girija S Sajjan; Yv Sashi Kanth
Journal:  J Conserv Dent       Date:  2013-09

7.  Evaluation of Two New Electronic Apex-Locator-Controlled Handpieces Using a NiTi Rotary File: An In Vitro Study.

Authors:  Kiumarz Nazarimoghadam; Hossein Labaf
Journal:  J Dent (Tehran)       Date:  2013-11-30
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.