Comparison of forced-air warming systems with upper body blankets using a copper manikin of the human body.

A Bräuer1, M J M English, N Steinmetz, N Lorenz, T Perl, U Braun, W Weyland.   

Abstract

BACKGROUND: Forced-air warming with upper body blankets has gained high acceptance as a measure for the prevention of intraoperative hypothermia. However, data on heat transfer with upper body blankets are not yet available. This study was conducted to determine the heat transfer efficacy of eight complete upper body warming systems and to gain more insight into the principles of forced-air warming.
METHODS: Heat transfer of forced-air warmers can be described as follows: Qdot;=h. DeltaT. A, where Qdot;= heat flux [W], h=heat exchange coefficient [W m-2 degrees C-1], DeltaT=temperature gradient between the blanket and surface [ degrees C], and A=covered area [m2]. We tested eight different forced-air warming systems: (1) Bair Hugger and upper body blanket (Augustine Medical Inc. Eden Prairie, MN); (2) Thermacare and upper body blanket (Gaymar Industries, Orchard Park, NY); (3) Thermacare (Gaymar Industries) with reusable Optisan upper body blanket (Willy Rüsch AG, Kernen, Germany); (4) WarmAir and upper body blanket (Cincinnati Sub-Zero Products, Cincinnati, OH); (5) Warm-Gard and single use upper body blanket (Luis Gibeck AB, Upplands Väsby, Sweden); (6) Warm-Gard and reusable upper body blanket (Luis Gibeck AB); (7) WarmTouch and CareDrape upper body blanket (Mallinckrodt Medical Inc., St. Luis, MO); and (8) WarmTouch and reusable MultiCover trade mark upper body blanket (Mallinckrodt Medical Inc.) on a previously validated copper manikin of the human body. Heat flux and surface temperature were measured with 11 calibrated heat flux transducers. Blanket temperature was measured using 11 thermocouples. The temperature gradient between the blanket and surface (DeltaT) was varied between -8 and +8 degrees C, and h was determined by linear regression analysis as the slope of DeltaT vs. heat flux. Mean DeltaT was determined for surface temperatures between 36 and 38 degrees C, as similar mean skin surface temperatures have been found in volunteers. The covered area was estimated to be 0.35 m2.
RESULTS: Total heat flow from the blanket to the manikin was different for surface temperatures between 36 and 38 degrees C. At a surface temperature of 36 degrees C the heat flows were higher (4-26.6 W) than at surface temperatures of 38 degrees C (2.6-18.1 W). The highest total heat flow was delivered by the WarmTouch trade mark system with the CareDrape trade mark upper body blanket (18.1-26.6 W). The lowest total heat flow was delivered by the Warm-Gard system with the single use upper body blanket (2.6-4 W). The heat exchange coefficient varied between 15.1 and 36.2 W m-2 degrees C-1, and mean DeltaT varied between 0.5 and 3.3 degrees C.
CONCLUSION: We found total heat flows of 2.6-26.6 W by forced-air warming systems with upper body blankets. However, the changes in heat balance by forced-air warming systems with upper body blankets are larger, as these systems are not only transferring heat to the body but are also reducing heat losses from the covered area to zero. Converting heat losses of approximately 37.8 W to heat gain, results in a 40.4-64.4 W change in heat balance. The differences between the systems result from different heat exchange coefficients and different mean temperature gradients. However, the combination of a high heat exchange coefficient with a high mean temperature gradient is rare. This fact offers some possibility to improve these systems.

Entities:  

Mesh:

Year:  2002        PMID: 12190797     DOI: 10.1034/j.1399-6576.2002.460807.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  4 in total

1.  Assessment of the efficiency of warming devices during neonatal surgery.

Authors:  Philippe Buisson; Véronique Bach; Elmountacer Billah Elabbassi; Karen Chardon; Stéphane Delanaud; Jean-Pierre Canarelli; Jean-Pierre Libert
Journal:  Eur J Appl Physiol       Date:  2004-09       Impact factor: 3.078

2.  Intensified thermal management for patients undergoing transcatheter aortic valve implantation (TAVI).

Authors:  Ivo F Brandes; Marc Jipp; Aron F Popov; Ralf Seipelt; Michael Quintel; Anselm Bräuer
Journal:  J Cardiothorac Surg       Date:  2011-09-25       Impact factor: 1.637

3.  Comparison of upper and lower body forced air blanket to prevent perioperative hypothermia in patients who underwent spinal surgery in prone position: a randomized controlled trial.

Authors:  Jae Hwa Yoo; Si Young Ok; Sang Ho Kim; Ji Won Chung; Sun Young Park; Mun Gyu Kim; Ho Bum Cho; Sang Hoon Song; Yun Jeong Choi; Hyun Ju Kim; Hong Chul Oh
Journal:  Korean J Anesthesiol       Date:  2021-05-13

4.  Efficacy of active forced air warming during induction of anesthesia to prevent inadvertent perioperative hypothermia in intraoperative warming patients: Comparison with passive warming, a randomized controlled trial.

Authors:  Jae Hwa Yoo; Si Young Ok; Sang Ho Kim; Ji Won Chung; Sun Young Park; Mun Gyu Kim; Ho Bum Cho; Sang Hoon Song; Chae Yeon Cho; Hong Chul Oh
Journal:  Medicine (Baltimore)       Date:  2021-03-26       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.