Blood values of juvenile northern elephant seals (Mirounga angustirostris) obtained using a portable clinical analyzer.

R Scott Larsen1, Martin Haulena, Carol B Grindem, Frances M D Gulland.   

Abstract

BACKGROUND: Sick, injured, or orphaned juvenile northern elephant seals (Mirounga angustisrostris) treated at rehabilitation centers frequently present with abnormalities in blood sodium, potassium, chloride, BUN, and glucose concentrations, and HCT. These abnormalities could be detected rapidly using a portable blood analyzer, but results with this analysis method do not necessarily equate with those obtained using other techniques.
OBJECTIVE: The objective of this study was to better assess the clinical relevance of values obtained from a portable analyzer and to compare the results with values obtained using more common methods of analysis.
METHODS: Heparinized whole blood samples were collected from 20 rehabilitated juvenile northern elephant seals. A portable clinical analyzer (i-STAT, i-STAT Corp, East Windsor, NJ, USA) was used to establish baseline values. Serum biochemical values were obtained using an automated chemical analyzer (Olympus AU5200, Olympus America, Melville, NY, USA). HCT was determined using EDTA whole blood and a cell counter.
RESULTS: Using the portable analyzer, mean (minimum-maximum) values were obtained for sodium, 143 (132-146) mmol/L; potassium, 4.4 (3.9-5.8) mmol/L; chloride, 106 (101-109) mmol/L; BUN, 1.8 (1.1-2.4) mmol/L; glucose, 7.55 (5.99-8.49) mmol/L; and HCT, 0.55 (0.52-0.61) L/L. Average differences between methods were small for potassium (-0.45 mmol/L), BUN (0.1 mmol/L), and HCT (0.037 L/L) but were large for sodium (-6.8 mmol/L), chloride (-6.4 mmol/L), and glucose (-0.56 mmol/L).
CONCLUSIONS: These results suggest that the i-STAT portable analyzer could be useful for clinically assessing juvenile elephant seals. However, when making medical decisions, the clinician should be aware of differences associated with various analyzers and sample types.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12189595     DOI: 10.1111/j.1939-165x.2002.tb00288.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  3 in total

1.  Obtaining accurate glucose measurements from wild animals under field conditions: comparing a hand held glucometer with a standard laboratory technique in grey seals.

Authors:  Kimberley A Bennett; Lucy M Turner; Sebastian Millward; Simon E W Moss; Ailsa J Hall
Journal:  Conserv Physiol       Date:  2017-02-27       Impact factor: 3.079

2.  Use of portable blood physiology point-of-care devices for basic and applied research on vertebrates: a review.

Authors:  Lauren J Stoot; Nicholas A Cairns; Felicia Cull; Jessica J Taylor; Jennifer D Jeffrey; Félix Morin; John W Mandelman; Timothy D Clark; Steven J Cooke
Journal:  Conserv Physiol       Date:  2014-04-04       Impact factor: 3.079

3.  Validation of the i-STAT and HemoCue systems for the analysis of blood parameters in the bar-headed goose, Anser indicus.

Authors:  T S Harter; M Reichert; C J Brauner; W K Milsom
Journal:  Conserv Physiol       Date:  2015-05-19       Impact factor: 3.079

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.