Optimum iodine concentration of contrast material through microcatheters: hydrodynamic analysis of experimental results.

Teiyu Yamauchi1, Shigeru Furui, Arimi Harasawa, Makoto Ishimura, Tatsuhiko Imai, Toshimitsu Hayashi.   

Abstract

It is important to increase the iodine delivery rate (I), that is the iodine concentration of the contrast material (C) x the flow rate of the contrast material (Q), through a catheter to obtain high quality arteriograms. The iodine delivery rate varies depending on the iodine concentration of the contrast material. The purpose of this study is to estimate the optimum iodine concentration (Copt) of contrast material (i.e. the iodine concentration at which I becomes maximum) through a microcatheter of a given length (L), inner diameter (D) and injection pressure (P). Iohexol, ioversol and iopamidol of 11-12 iodine concentrations (140-350, 160-350 or 160-370 mg cm(-3)) at 37 degrees C are used. I and Reynolds number (Re) of the flow of each contrast material through four microcatheters (0.49-0.68 mm in inner diameter, 1000-1500 mm in length) at injection pressures of 1.38, 2.76, 4.14 and 5.52 x 10(6) Pa (200, 400, 600 and 800 pounds per square inch) are obtained experimentally. The relationships between I and C and between I and Re are examined for each catheter and injection pressure. Copt is 160-280 mg cm(-3) for iohexol, 180-280 mg cm(-3) for ioversol and 200-300 mg cm(-3) for iopamidol. I becomes maximum when Re approximates the critical Reynolds number (Re approximately 2300). Utilizing this principle, we can estimate Copt and its flow rate through a microcatheter of a given L, D and P.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12171337     DOI: 10.1088/0031-9155/47/14/310

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  A high-precision contrast injector for small animal x-ray digital subtraction angiography.

Authors:  Ming de Lin; Lutao Ning; Cristian T Badea; Nilesh N Mistry; Yi Qi; G Allan Johnson
Journal:  IEEE Trans Biomed Eng       Date:  2008-03       Impact factor: 4.538

2.  Minimally Invasive Monitoring of Chronic Central Venous Catheter Patency in Mice Using Digital Subtraction Angiography (DSA).

Authors:  Giovanna Figueiredo; Teresa Fiebig; Stefanie Kirschner; Omid Nikoubashman; Lisa Kabelitz; Ahmed Othman; Andrea Nonn; Martin Kramer; Marc A Brockmann
Journal:  PLoS One       Date:  2015-06-22       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.