Abstract

While the hemodynamic and clinical significance of automatic mode switching (AMS) in patients with pacemakers has been demonstrated, the clinical behavior of AMS algorithms differ widely according to the manufacturers and pacemaker models. In general, a "rate-cutoff" detection method of atrial tachyarrhythmias provides a rapid AMS onset and resynchronization to sinus rhythm at the termination of atrial tachyarrhythmias, but may cause intermittent oscillations between the atrial tracking and AMS mode. This can be minimized with a "counter" of total number of high rate events before the AMS occurs. The use of a "running average" algorithm results in more stable rate control during AMS by reducing the incidence of oscillations, but at the expense of delayed AMS onset and resynchronization to sinus rhythm. Algorithms may be combined to fine tune the AMS response and to avoid rapid fluctuation in pacing rate. Appropriate programming of atrial sensitivity, and the avoidance of ventriculoatrial cross-talk are essential for optimal AMS performance.

Entities:  

Mesh:

Year:  2002        PMID: 12164453     DOI: 10.1046/j.1460-9592.2002.01094.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

Review 1.  Follow up and optimisation of cardiac pacing.

Authors:  Paul R Roberts
Journal:  Heart       Date:  2005-09       Impact factor: 5.994

2.  Shortening of the sensed AV delay of a dual chamber pacemaker during normal sinus rhythm.

Authors:  S Serge Barold; Chu-Pak Lau; Hung-Fat Tse; Michael Hess
Journal:  J Interv Card Electrophysiol       Date:  2003-08       Impact factor: 1.900

3.  Automatic mode switching in atrial fibrillation.

Authors:  Giuseppe Stabile; Antonio De Simone; Enrico Romano
Journal:  Indian Pacing Electrophysiol J       Date:  2005-07-01
  3 in total