Absorbed dose behind eye shields during kilovoltage photon radiotherapy.

C R Baker1, F Luhana, S J Thomas.   

Abstract

The absorbed dose at the position of the lens of the eye under lead or tungsten eye shields during kilovoltage photon radiotherapy is critically dependent not so much on the thickness of the eye shield itself as on the size of the treatment field and the diameter of the shield used. Whilst dose from primary photons is easily attenuated to relatively insignificant levels by a few millimetres of lead or tungsten, scattered photons from outside the shielded area can provide over 25% of the prescribed dose. Since backscatter factors do not increase monotonically with photon energy, it is not safe to assume that the highest photon energy used will provide the highest dose. A simple method to estimate the dose under an eye shield based on tabulated backscatter factors is shown. Measurements under commercially available eye shields were made to verify the expression and to determine the attenuation of primary photons. Predicted and measured absorbed dose under the eye shields were found to agree to within 1% of the prescribed dose. The relative dose due to primary photons beneath the eye shields was found to be less than 0.1% and 0.5 (+/-0.1)% for the 150 kV and 260 kV beams, respectively. This is considerably less than the dose from backscattered radiation.

Mesh:

Substances:

Year:  2002        PMID: 12153943     DOI: 10.1259/bjr.75.896.750685

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  2 in total

1.  Calculation of excess dose to the eye phantom due to a distanced shielding for electron therapy in head and neck cancers.

Authors:  Keyvan Jabbari; Mahnaz Roayaei; Hosein Saberi
Journal:  J Med Signals Sens       Date:  2012-07

2.  Application of a dummy eye shield for electron treatment planning.

Authors:  Sei-Kwon Kang; Soah Park; Taejin Hwang; Kwang-Ho Cheong; Taejin Han; Haeyoung Kim; Me-Yeon Lee; Kyoung Ju Kim; Do Hoon Oh; Hoonsik Bae
Journal:  J Radiat Res       Date:  2012-08-21       Impact factor: 2.724

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.