Performance characteristics of an automated high-sensitivity C-reactive protein assay on the Dimension RXL analyzer.

Barun K De1, Linda G Smith, William E Owen, William L Roberts.   

Abstract

BACKGROUND: C-reactive protein (CRP) is a nonspecific marker of inflammation that can be used as a marker for atherosclerotic risk. This application requires increased precision at low CRP concentrations compared to traditional assays.
METHODS: The Dimension RXL is an automated chemistry analyzer for central laboratory use. The limit of detection, limit of quantification, linearity and imprecision of a high-sensitivity CRP assay developed for it were assessed. Method comparison studies were performed using samples both inside and outside the reference interval. The presence of a prozone effect was also evaluated.
RESULTS: The limit of detection was 0.7 mg/l. The method was linear from 2 to 60 mg/l and from 1 to 60 mg/l using systematic error limits of 10% and 20%, respectively. The total imprecision was <10% for CRP concentrations above 1.5 mg/l. No prozone effect was seen at a CRP concentration of 450 mg/l, the highest concentration tested. Using samples from 212 apparently healthy adults, the Dimension RXL method demonstrated good concordance with the BN II high-sensitivity CRP method for samples in the highest quartile. It also compared well using samples with elevated CRP concentrations.
CONCLUSIONS: The Dimension RXL high-sensitivity CRP method may be adequate for atherosclerotic risk prediction in clinical practice if accurate and precise measurement is only required for the highest quartile. However, the total error of this method for CRP concentrations <3 mg/l appears too large for accurate assignment to lower risk groups.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12135817     DOI: 10.1016/s0009-8981(02)00185-7

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  4 in total

1.  Diagnostic accuracy and clinical role of rapid C-reactive protein testing in HIV-infected individuals with presumed tuberculosis in South Africa.

Authors:  P K Drain; L Mayeza; P Bartman; R Hurtado; P Moodley; S Varghese; G Maartens; G G Alvarez; D Wilson
Journal:  Int J Tuberc Lung Dis       Date:  2014-01       Impact factor: 2.373

2.  Antibodies against Mycobacterial proteins as biomarkers for HIV-associated smear-negative tuberculosis.

Authors:  Michael Siev; Douglas Wilson; Supreet Kainth; Victoria O Kasprowicz; Catherine M Feintuch; Elizabeth R Jenny-Avital; Jacqueline M Achkar
Journal:  Clin Vaccine Immunol       Date:  2014-03-26

3.  Measurement of C-reactive protein: two high sensitivity methods compared.

Authors:  Roberto Dominici; Paola Luraschi; Carlo Franzini
Journal:  J Clin Lab Anal       Date:  2004       Impact factor: 2.352

4.  Analytical performance of a highly sensitive C-reactive protein-based immunoassay and the effects of laboratory variables on levels of protein in blood.

Authors:  Najib Aziz; John L Fahey; Roger Detels; Anthony W Butch
Journal:  Clin Diagn Lab Immunol       Date:  2003-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.