Literature DB >> 12135549

Effects of one-plane and two-plane external fixation on sheep osteotomy healing and complications.

Gert D Krischak1, Andreas Janousek, Steffen Wolf, Peter Augat, Lothar Kinzl, Lutz E Claes.   

Abstract

OBJECTIVES: Comparison of one-plane and two-plane external fixation in terms of successful healing, incidence of complications, and biomechanical stability in a sheep model.
BACKGROUND: Rigid fixation is preferred in open and comminuted fractures with a reduced blood supply, preventing infection and healing delay, but more often a flexible device is recommended even in unfavourable healing conditions.
METHODS: The left tibiae of fifteen sheep were osteotomized and laterally fixed with a four-screw unilateral fixator frame (axial stiffness 183 N/mm) to a 3 mm gap size. In 9 of 15 sheep, an additional four-screw unilateral external fixator was anterolaterally attached (total axial stiffness of both frames 388 N/mm). After sacrificing, quality of osteotomy healing was assessed by mechanical and radiological evaluations. Osteogenesis was measured using fluorescence microscopy.
RESULTS: Two distal fractures through the pin-tracks, three non-unions and four deep infections occurred after two-plane fixation. These failures excluded, osteotomy healing showed inferior results after two-plane fixation with reduced callus formation, bone mineral content, and bending stiffness amounts, respectively. Osteogenesis was halved following two-plane fixation in the remaining sheep.
CONCLUSIONS: Two-plane fixation was not sufficient to reach successful osteotomy healing in our study. While higher rigidity was expected to prevent complications, healing in this group might have been disturbed by a reduced blood supply. The optimal stabilisation for a given fracture depends on many factors, including the biomechanical and biological environment. RELEVANCE: Considering our results and the literature discussed in this manuscript, good bone healing with minor risks of infections can be achieved using an unilateral one-plane fixator with only four screws, and its application on a muscle free position like the medial and anterior site of the sheep tibia.

Entities:  

Mesh:

Year:  2002        PMID: 12135549     DOI: 10.1016/s0268-0033(02)00039-6

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  Double-plating of ovine critical sized defects of the tibia: a low morbidity model enabling continuous in vivo monitoring of bone healing.

Authors:  Joachim A Hahn; Tanja S Witte; Daniel Arens; Alexandra Pearce; Simon Pearce
Journal:  BMC Musculoskelet Disord       Date:  2011-09-29       Impact factor: 2.362

2.  Biomechanical features of six design of the delta external fixator for treating Pilon fracture: a finite element study.

Authors:  Muhammad Hanif Ramlee; Mohd Ayub Sulong; Evelyn Garcia-Nieto; Daniel Angure Penaranda; Antonio Ros Felip; Mohammed Rafiq Abdul Kadir
Journal:  Med Biol Eng Comput       Date:  2018-04-21       Impact factor: 2.602

3.  Unicortical self-drilling external fixator pins reduce thermal effects during pin insertion.

Authors:  Markus Greinwald; Patrick A Varady; Peter Augat
Journal:  Eur J Trauma Emerg Surg       Date:  2017-12-14       Impact factor: 3.693

  3 in total