| Literature DB >> 12132948 |
Thomas R LaVoy1, Walter Huda, Kent M Ogden.
Abstract
The objectives of this study were to document imaging physics parameters associated with mammography physics surveys, and investigate how the choice of tube potential affects average glandular dose (AGD) and x-ray exposure time. Data from 60 mammography units were obtained pertaining to representative values of mAs, exposure time, half value layer, AGD and film density when acquiring phantom images. The survey of clinical systems showed that for a normal sized breast as represented by the mammography accreditation phantom, 60% of these units were operated at 25 kVp, and 33% at 26 kVp. Median exposure times were 1.14 s at 25 kVp and 0.73 s at 26 kVp. The median AGD was 1.62 mGy at 25 kVp and 1.51 mGy at 26 kVp. As expected, the choice of x-ray tube potential did not significantly affect the median film density value of 1.5. Five clinical systems, all from different vendors, had measurements performed of the AGD and x-ray exposure time as a function of x-ray tube potential at a constant film density. For a typical clinical x-ray unit, increasing the x-ray tube potential from 25 to 28 kVp reduced the exposure time by 50%, and reduced the AGD by 26%.Entities:
Mesh:
Year: 2002 PMID: 12132948 PMCID: PMC5724594 DOI: 10.1120/jacmp.v3i3.2572
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
List of manufacturers and screen‐film combinations for the 60 units included in this survey.
| Mammography Systems | Sterling | Kodak | Fuji | |||||
|---|---|---|---|---|---|---|---|---|
| Manufacture | Model | Microvision C | MIN‐R M | 2000 | E | AD | MA | Total |
| M‐II | 1 | 2 | 2 | |||||
| M‐II E | 1 | 2 | 3 | |||||
| LoRad | M‐III | 5 | 9 | 14 | ||||
| M‐IV | 2 | 6 | 1 | 9 | ||||
| TransPo | 1 | 1 | 2 | |||||
| General Electric | 800T | 1 | 1 | |||||
| DMR | 4 | 4 | ||||||
| Mammomat 2 | 1 | 4 | 1 | |||||
| Siemens | Mammomat | 3 | 4 | 2 | 1 | 10 | ||
| 3000 | ||||||||
| Instrumentarium | Alpha IQ | 3 | 2 | 3 | ||||
| Alpha RT | 2 | 2 | ||||||
| Bennett | MF‐150 | 1 | 1 | 2 | ||||
| Contour | 1 | 1 | ||||||
| Amerisys | Arisa M | 2 | 2 | |||||
| Transworld | MAM‐CP | 1 | 1 | 2 | ||||
| Philips | Diagnost UM | 1 | 1 | |||||
| Total | 19 | 2 | 30 | 4 | 3 | 2 | 60 | |
Variation of four mammography imaging parameters (mAs, time, HVL and AGD) with x‐ray tube potential (25 and 28 kVp).
| Manufacturer | Model | mAs | Time | HVL | AGD |
|---|---|---|---|---|---|
| Siemens | M‐3000 | 107.5/50.6 | 0.72/0.38 | 0.315/0.349 | 1.42/1.09 |
| GE | DMR | 129/54 | 1.29/0.54 | 0.334/0.367 | 1.83/1.29 |
| Instrumentarium | Alpha IQ | 93/45 | 0.93/0.53 | 0.336/0.372 | 1.51/1.19 |
| Lorad | M‐IV | 119/53.4 | 1.2/0.53 | 0.291/0.322 | 1.60/1.18 |
| Bennet | Contour | 185.6/92.6 | 1.24/0.62 | 0.277/0.315 | 1.71/1.20 |
First value is at 25 kVp/second value is at 28 kVp.
Statistical analysis of imaging performance characteristics found for mammography units operated at 25 and 26 kVp.
| mAs | Exposure time (s) | Half value layer (mm Al) | AGD (mGy) | Film density | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Percentile value | 25 kVp | 26 kVp | 25 kVp | 26 kVp | 25 kVp | 26 kVp | 25 kVp | 26 kVp | 25 kVp | 26 kVp |
| 10% | 80 | 72 | 0.84 | 0.51 | 0.293 | 0.310 | 1.39 | 1.26 | 1.39 | 1.42 |
| 30% | 87 | 77 | 1.03 | 0.57 | 0.299 | 0.328 | 1.54 | 1.37 | 1.46 | 1.48 |
| 50% | 95 | 84 | 1.14 | 0.73 | 0.309 | 0.331 | 1.62 | 1.51 | 1.51 | 1.52 |
| 70% | 106 | 93 | 1.23 | 1.05 | 0.319 | 0.334 | 1.70 | 1.64 | 1.53 | 1.53 |
| 90% | 122 | 98 | 1.39 | 1.24 | 0.327 | 0.341 | 1.81 | 1.75 | 1.55 | 1.56 |
Figure 1Average glandular dose versus measured exposure time for a breast with a thickness equivalent to the ACR accreditation phantom, and a 50% glandularity.
Figure 2Relative change in radiographic parameters as a function of kVp: (a) mAs, (b) time, (c) half value layer, and (d) average glandular dose.
Summary of least square fit data for the equation (parameter and the corresponding coefficient of determination , given by the solid lines in Fig. 2.
| Parameter | α | β | γ |
| |
|---|---|---|---|---|---|
| mAs | 0.04125 | –2.361 | 34.24 | 0.996 | |
| time | 0.03925 | –2.246 | 32.61 | 0.995 | |
| HVL |
| 0.0898 | –0.6192 | 0.998 | |
| AGD |
| –0.6050 | 10.03 | 0.998 |
Relative parameter values as a function of kVp, normalized to 100% at 25 kVp. These data are obtained using the least square fit parameter given in Table III.
| Relative parameter value (normalized to 100% at 25 kVp) | ||||
|---|---|---|---|---|
| kVp | mAs | Time | HVL | AGD |
| 25 | 99.4% | 99.1%100% | 100.1% | |
| 26 | 73.3% | 74.6% | 104.0% | 89.2% |
| 27 | 56.5% | 58.7% | 107.6% | 80.7% |
| 28 | 47.0% | 49.7% | 111.1% | 73.7% |