The calibration of a liquid I-125 source in an older standard dose calibration system is presented. The calibration factor agrees well with factors established by the NIST for newer dose calibration systems. The determination of the source activity is necessary to accurately calculate the time required to deliver the prescribed dose.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12132943 PMCID: PMC5724600 DOI: 10.1120/jacmp.v3i3.2565
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Dial setting for the calibration of a liquid source.
| Device | Dial setting recommendation | Source of recommendation |
|---|---|---|
| CRC5 | 475 | Measurement |
| CRC‐12 | 497 | NIST Report |
| CRC‐35R | 469 | NIST Report |