Commissioning and periodic quality assurance of a clinical electronic portal imaging device.

D A Low1, E E Klein, D K Maag, W E Umfleet, J A Purdy.   

Abstract

PURPOSE: An electronic portal imaging device (EPID) was recently installed on our dual-energy linear accelerator. Commissioning and quality assurance techniques were developed for the EPID. METHODS AND MATERIALS: A commissioning procedure was developed consisting of five parts: (a) physical operation and safety; (b) image acquisition, resolution, and sensitivity calibration; (c) image storage, analysis, and handling; (d) reference image acquisition; and (e) clinical operations.
RESULTS: The physical operation and safety tests relate to the motions of the unit, stability of the unit supports, safety interlocks, and interlock overrides. Imager contrast and spatial resolutions are monitored by imaging a contrast-detail phantom. The imager calibration procedure consists of a no-radiation image to compensate for signal offsets, as well as a "flat-field image." The flat-field image is taken with 5.0 cm of homogeneous phantom material placed at isocenter to provide some photon scatter and to approximate the presence of a patient. Daily quality assurance procedures consists of safety tests and the acquisition and inspection of images of the contrast-detail phantom. After 1 year, the frequency of the daily procedure was reduced to weekly. Quarterly QA procedures are conducted by the physicist and consist of the same procedures conducted in the weekly test. The annual QA procedure consists of a duplication of the commissioning procedure.
CONCLUSION: The procedures discussed in this article were applied to an ionization-chamber device. They have been useful in identifying difficulties with the EPID operation, including the need for recalibrating and monitoring the accelerator output stability.

Entities:  

Mesh:

Year:  1996        PMID: 12118539     DOI: 10.1016/0360-3016(95)02096-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Quality assurance for kilo- and megavoltage in-room imaging and localization for off- and online setup error correction.

Authors:  James M Balter; Larry E Antonuk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

2.  A quality assurance phantom for electronic portal imaging devices.

Authors:  Indra J Das; Minsong Cao; Chee-Wai Cheng; Vladimir Misic; Klaus Scheuring; Edmund Schüle; Peter A S Johnstone
Journal:  J Appl Clin Med Phys       Date:  2011-02-02       Impact factor: 2.102

3.  Commissioning of a dedicated commercial Co-60 total body irradiation unit.

Authors:  Jay Burmeister; Adrian Nalichowski; Michael Snyder; Robert Halford; Geoff Baran; Brian Loughery; Ahmad Hammoud; Joe Rakowski; Todd Bossenberger
Journal:  J Appl Clin Med Phys       Date:  2018-03-11       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.