Cardiac pacemaker induced ventricular fibrillation during surgical diathermy.

H J Orland, D Jones.   

Abstract

Two demand pacemakers were observed to repeatedly fibrillate the heart during clinical surgery employing diathermy. Marked hypotension associated with a rapid, irregular venticular action was observed in another patient who had a fixed rate pacemaker implanted. Pacemakers were tested with function generators over a range of frequencies and some increased their rate to between 200-600 impulse/min. Animal experiments confirmed that, contrary to manufacturers advice, surgical diathermy produces myocardial fibrillation in the presence of some pacemakers. Mechanisms of these effects and their clinical significance are discussed.

Entities:  

Mesh:

Year:  1975        PMID: 1211612     DOI: 10.1177/0310057X7500300407

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  1 in total

1.  Transducer offset by electrocautery resulting in erroneous blood pressure measurement.

Authors:  B Milne; F W Cervenko; M B Henderson; P J Westra
Journal:  Can Anaesth Soc J       Date:  1986-03