| Literature DB >> 12113980 |
Koen De Cremer1, Rita Cornelis, Karel Strijckmans, Richard Dams, Norbert Lameire, Raymond Vanholder.
Abstract
The behaviour of free [48V]vanadate and [48V]vanadium-transferrin complex was investigated on five different anion-exchange columns (Mono Q 5/5 HR, Hitrap Q HP, Sepharose Q FF, Sepharose DEAE FF and Hitrap Q XL). The recovery of both V-compounds was quantitative. The peak shape and retention time of vanadate varied according to the type of column. The vanadium-transferrin complex also showed different elution patterns depending on the type of column. Especially in case of the Sepharose Q FF, Mono Q 5/5 HR and Hitrap Q XL columns the vanadium-transferrin binding was degraded during elution on the column. The results clearly prove that care should be taken as to the choice of column for speciation purposes of vanadium compounds in order to prevent various artefacts showing up in the chromatograms. A Hitrap Q HP column was used to fractionate different vanadium compounds in rat serum. Copyright 2002 Elsevier Science BV.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12113980 DOI: 10.1016/s1570-0232(02)00278-7
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205