Laboratory testing of hydrocephalus shunts -- conclusion of the U.K. Shunt evaluation programme.

Z Czosnyka1, M Czosnyka, H K Richards, J D Pickard.   

Abstract

16 models of valves, currently in use in the U.K., have been tested long-term in the U.K. Shunt Evaluation Laboratory according to the protocol based on the new ISO 7197 standard. Valves tested were: Medtronic PS Medical: Delta Valve, Flow Control and Lumbo-Peritoneal Shunt, Heyer-Schulte Nero-Care: In-line, Low Profile and Pudenz Flushing Valve, Codman: Codman-Hakim Programmable, Hakim Precision, Accu-Flo, Holter, Uni-Shunt, and siphon-preventing device -- SiphonGuard, NMT: Orbis-Sigma Valve, Omni-Shunt and Hakim Valve, Sophysa: Sophy Programmable Valve, Radionics: Contour-Flex Valve. The majority of the valves had a non-physiologically low hydrodynamic resistance (with the exception of Orbis-Sigma, PS Lumbo-Peritoneal and Heyer-Schulte In-Line). This may result in overdrainage both related to posture and during nocturnal cerebral vasogenic waves. A long distal catheter increases the resistance of these valves by 100-200%. Drainage through valves without siphon-preventing mechanism is very sensitive to body posture. This may produce grossly negative intracranial pressure after implantation. A few shunts (Delta, Low Profile and Pudenz-Flushing with Anti-Siphon Devices) offer a reasonable resistance to negative outlet pressure, and hence potentially might prevent complications related to overdrainage. On the other hand, valves with siphon-preventing devices may be blocked by raised subcutaneous pressure (exception: SiphonGuard, but this device may block the drainage because of its faulty design). In most of the silicone-diaphragm valves, closing pressure varied and reached values lower than that specified by the manufacturer (exception: Heyer-Schulte Pudenz Flushing Valve). All programmable valves are susceptible to overdrainage in the upright body position. Programmed settings may be changed by external magnetic fields. Most shunts are very sensitive to the presence of small particles in the drained fluid. The behavior of a valve revealed during such testing is of immediate relevance to the surgeon and may not be adequately described in the manufacturer's product information. These results are also relevant to the assessment of shunt function in-vivo using an infusion test.

Entities:  

Mesh:

Year:  2002        PMID: 12111485     DOI: 10.1007/s00701-002-0922-9

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  7 in total

1.  [Gravity valves for idiopathic normal pressure hydrocephalus. A Prospective study of 60 patients].

Authors:  U Meier
Journal:  Nervenarzt       Date:  2004-06       Impact factor: 1.214

2.  Features of the Sinushunt and its influence on the cerebrospinal fluid system.

Authors:  A Eklund; L-O D Koskinen; J Malm
Journal:  J Neurol Neurosurg Psychiatry       Date:  2004-08       Impact factor: 10.154

3.  A mathematical model of blood, cerebrospinal fluid and brain dynamics.

Authors:  Andreas A Linninger; Michalis Xenos; Brian Sweetman; Sukruti Ponkshe; Xiaodong Guo; Richard Penn
Journal:  J Math Biol       Date:  2009-02-15       Impact factor: 2.259

4.  Surgical outcome of the shunt: 15-year experience in a single institution.

Authors:  Sara Iglesias; Bienvenido Ros; Álvaro Martín; Antonio Carrasco; Miguel Segura; Andrea Delgado; Francisca Rius; Miguel Ángel Arráez
Journal:  Childs Nerv Syst       Date:  2016-08-05       Impact factor: 1.475

5.  Infusion technique can be used to distinguish between dysfunction of a hydrocephalus shunt system and a progressive dementia.

Authors:  A Eklund; B Lundkvist; L-O D Koskinen; J Malm
Journal:  Med Biol Eng Comput       Date:  2004-09       Impact factor: 2.602

6.  Shunt technology for infants and a lifetime.

Authors:  Víctor J Fernández Cornejo; Samer K Elbabaa
Journal:  Childs Nerv Syst       Date:  2021-07-08       Impact factor: 1.475

7.  In vitro hydrodynamic properties of the Miethke ProGAV hydrocephalus shunt.

Authors:  David M Allin; Zofia H Czosnyka; Marek Czosnyka; Hugh K Richards; John D Pickard
Journal:  Cerebrospinal Fluid Res       Date:  2006-06-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.