R Ficarra1, M L Calabrò, P Cutroneo, S Tommasini, S Melardi, M Semreen, S Furlanetto, P Ficarra, G Altavilla.
Abstract
A rapid and sensitive RP-HPLC method with UV detection for routine control of oxaliplatin in a pharmaceutical formulation (Eloxatin) was developed. Quantitation was accomplished with the internal standard method. The procedure was validated by linearity (correlation coefficient=0.999948), accuracy, robustness and intermediate precision. Experimental design was used during validation to calculate method robustness and intermediate precision. For robustness test three factors were considered: percentage v/v of acetonitrile, flow rate and temperature; an increase in the flow rate results in a decrease of the drug found concentration, while the percentage of organic modifier and temperature have no important effect on the response. For intermediate precision measure the considered variables were: analyst, equipment and days. The RSD value (2.27%, n=24) indicated a good precision of the analytical method.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12110395 DOI: 10.1016/s0731-7085(02)00151-6
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935