Literature DB >> 12094991

Comparison of characteristics of photon and electron beams generated by Philips/Elekta and Varian linear accelerators.

Haijun Song1, Ying Xiao, James M Galvin.   

Abstract

Commissioning data of the common energy of 6 MV photon beams are analyzed for seven Philips/ Elekta linacs of five different models: SL 75/5, SL 15, SL 25, SLi Precise, and SLi. The observed beam quality %dd(10)x of 67.7% +/- 0.3 agrees with the manufacturer's specification of 67.5% +/- 2% but with a much smaller variation, suggesting the possibility of a more accurate beam quality guideline for acceptance commissioning of these linac models. Field size factors are less uniform, especially for the biggest field size of 40 x 40 cm2, but all agree within 1.1% between the five models. These 6 MV photon beams are also shown to be 0.4 MV harder than the Varian 2100C model. For commissioning electron beams generated by the Philips/Elekta models SL 15, SLi Precise and SLi, guidelines can be found from this work for the following parameters: actual field size at 100 SSD (up to +/- 0.6%) (different from manufacturer specified values at 95 SSD), and virtual source distance (up to +/- 0.7%). Significant differences are seen for R50 (up to 3.7 mm), sigma-theta-x, effective source distance, and cone size factor. However, except for R50 where Varian 2100C datasets are not available for comparison, the differences between the different Philips/Elekta models are comparable to those between the same Varian 2100C model linacs measured by us and Watts.

Entities:  

Mesh:

Year:  2002        PMID: 12094991     DOI: 10.1118/1.1477232

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Evaluation of an electron Monte Carlo dose calculation algorithm for electron beam.

Authors:  Ye Angela Hu; Haijun Song; Zhe Chen; Sumin Zhou; Fang-Fang Yin
Journal:  J Appl Clin Med Phys       Date:  2008-06-23       Impact factor: 2.102

2.  Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database.

Authors:  Christopher F Njeh
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

3.  Clinical implementation of electron energy changes of varian linear accelerators.

Authors:  Sean Zhang; Praimakorn Liengsawangwong; Patricia Lindsay; Karl Prado; Tzouh-Liang Sun; Roy Steadham; Xiaochun Wang; Mohammad Salehpour; Michael Gillin
Journal:  J Appl Clin Med Phys       Date:  2009-10-27       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.