Literature DB >> 12094973

Benchmarking of the dose planning method (DPM) Monte Carlo code using electron beams from a racetrack microtron.

Indrin J Chetty1, Jean M Moran, Daniel L McShan, Benedick A Fraass, Scott J Wilderman, Alex F Bielajew.   

Abstract

A comprehensive set of measurements and calculations has been conducted to investigate the accuracy of the Dose Planning Method (DPM) Monte Carlo code for dose calculations from 10 and 50 MeV scanned electron beams produced from a racetrack microtron. Central axis depth dose measurements and a series of profile scans at various depths were acquired in a water phantom using a Scanditronix type RK ion chamber. Source spatial distributions for the Monte Carlo calculations were reconstructed from in-air ion chamber measurements carried out across the two-dimensional beam profile at 100 cm downstream from the source. The in-air spatial distributions were found to have full width at half maximum of 4.7 and 1.3 cm, at 100 cm from the source, for the 10 and 50 MeV beams, respectively. Energy spectra for the 10 and 50 MeV beams were determined by simulating the components of the microtron treatment head using the code MCNP4B. DPM calculations are on average within +/- 2% agreement with measurement for all depth dose and profile comparisons conducted in this study. The accuracy of the DPM code illustrated in this work suggests that DPM may be used as a valuable tool for electron beam dose calculations.

Mesh:

Substances:

Year:  2002        PMID: 12094973     DOI: 10.1118/1.1481512

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Method for Fast CT/SPECT-Based 3D Monte Carlo Absorbed Dose Computations in Internal Emitter Therapy.

Authors:  S J Wilderman; Y K Dewaraja
Journal:  IEEE Trans Nucl Sci       Date:  2007-02-17       Impact factor: 1.679

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.