OBJECTIVE: Following the recommendation of the United State National Institute of Health Consensus Conference in 1993, otoacoustic emissions (OAE) are now used internationally for hearing screening. The GSI 70 OAE screener provides the means for carrying out OAE recordings within a short period of time and includes an automatic evaluation of results. The aim of this study was to determine the reliability of recordings in comparison with available standards in brainstem audiometry.
METHOD: OAE recorded in 29 hearing-impaired suspected infants and young children (aged 1 months-7 years old) in order to compare the result of the GSI 70 screener with the result of ABR. This study was conducted in the outpatient clinic of the Tokyo University Hospital. RESULT: Our study showed that the GSI 70 screener has higher false negative rate compared with ABR results (P<0.01). Consequently, an OAE measuring method, is also provides high sensitivity and easy to use. However, there was no trend toward increased refer rates with increased age.
CONCLUSION: Our findings show that the GSI 70 screener can meet the demands of systematic hearing screening in infants and young children, although there is a tendency to miss cochlear impaired cases.

Entities:  

Mesh:

Year:  2002        PMID: 12090949     DOI: 10.1016/s0165-5876(02)00071-x

Source DB:  PubMed          Journal:  Int J Pediatr Otorhinolaryngol        ISSN: 0165-5876            Impact factor:   1.675


  1 in total

1.  Study of the hearing in children born from pregnant women exposed to occupational noise: assessment by distortion product otoacoustic emissions.

Authors:  Eduardo Bezerra Rocha; Marisa Frasson de Azevedo; João Aragão Ximenes Filho
Journal:  Braz J Otorhinolaryngol       Date:  2007 May-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.