Marilyn S Hamilton1, David M Abel, Yolanda J Ballam, Mary K Otto, Angela F Nickell, Lisa M Pence, James R Appleman, Craig D Shimasaki, Komandoor E Achyuthan.
Abstract
Exploiting the high sensitivity of the chemiluminescence phenomenon, an accurate and sensitive point-of-care test, called the ZstatFlu-II test (ZymeTx, Inc., Oklahoma City, Okla.), was developed to detect influenza virus infections. The ZstatFlu-II test takes 20 min and requires approximately 2 min of "hands-on" time for operational steps. The ZstatFlu-II test does not distinguish between infections with influenza virus types A and B. ZstatFlu-II test results are printed on Polaroid High-Speed Detector Film, allowing test results to be archived. A prototype version of the ZstatFlu-II test was evaluated during the 2000-to-2001 flu season with 300 nasal aspirate specimens from children at a pediatric hospital. Compared to culture, the ZstatFlu-II test had 88% sensitivity and 92% specificity. The Directigen test had a sensitivity of 75% and a specificity of 93%. The sensitivity of the ZstatFlu-II test was significantly higher than that of the Directigen test (P < 0.0574).Entities:
Mesh:
Year: 2002 PMID: 12089243 PMCID: PMC120541 DOI: 10.1128/JCM.40.7.2331-2334.2002