Performance of a revised cardiac troponin method that minimizes interferences from heterophilic antibodies.

Wesley J Kim1, Omar F Laterza, Karl G Hock, James F Pierson-Perry, Delores M Kaminski, Martine Mesguich, François Braconnier, Régine Zimmermann, Martina Zaninotto, Mario Plebani, Atef Hanna, George S Cembrowski, Mitchell G Scott.   

Abstract

BACKGROUND: Recent guidelines for use of cardiac troponin to detect cardiac damage and for cardiovascular risk stratification have made increasingly sensitive troponin assays important. Troponin assays continue to be plagued by interferences caused by heterophilic antibodies (HAs). We evaluated the performance of a revised cardiac troponin I (cTnI) assay designed to have increased analytical sensitivity and to minimize the effect of HAs.
METHODS: The revised Dade Behring Dimension cTnI assay was evaluated according to NCCLS EP5-A at five institutions. Plasma samples from 14 309 patients were assayed by the original Dimension cTnI assay. To identify samples that may have interfering HAs, samples with values >1.4 microg/L were reanalyzed on the Dade Behring Stratus CS cTnI assay. Samples with possible interfering antibodies were also analyzed before and after selective absorbance studies on the revised Dade Behring Dimension cTnI assay.
RESULTS: The limit of quantification in the revised method was 0.1 microg/L with imprecision (CV) of 11-17% at 0.1 microg/L. Values correlated well with the Stratus CS cTnI method: revised = 1.06(original) + 0.01; r = 0.98, S(y/x) = 0.25 microg/L). Falsely increased results consistent with myocardial infarction by the original Dimension cTnI assay and presumably attributable to HAs were identified in 0.17% of all patients with samples submitted for cTnI analysis. The revised Dimension cTnI assay eliminated the interference in 17 of 25 samples identified and greatly decreased the interference in the other 8.
CONCLUSIONS: The revised Dimension cTnI method greatly minimizes the effect of interfering HAs. It also exhibits analytical performance characteristics consistent with recent guidelines for use of this assay to detect cardiac damage.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12089171

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  4 in total

1.  Postoperative myocardial injury after major head and neck cancer surgery.

Authors:  Peter Nagele; Lesley K Rao; Mrudula Penta; Dorina Kallogjeri; Edward L Spitznagel; Laura F Cavallone; Brian Nussenbaum; Jay F Piccirillo
Journal:  Head Neck       Date:  2010-09-30       Impact factor: 3.147

2.  Heterophile antibodies, false-positive troponin, and acute coronary syndrome: a case report indicating a pitfall in clinical practice.

Authors:  Nenad Lakusic; Ivana Sopek Merkas; Daren Lucinger; Darija Mahovic
Journal:  Eur Heart J Case Rep       Date:  2021-02-04

Review 3.  On the Effect of Heterophilic Antibodies on Serum Levels of Cardiac Troponins: A Brief Descriptive Review.

Authors:  Aleksey Michailovich Chaulin
Journal:  Life (Basel)       Date:  2022-07-24

4.  Recommendations on Use of Biochemical Markers in Acute Coronary Syndrome: IFCC Proposals.

Authors:  Mauro Panteghini
Journal:  EJIFCC       Date:  2003-07-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.