Continuous pH monitoring using the Paratrend 7 inserted into a peripheral vein in a patient with shock and congenital lactic acidosis.

R Blaine Easley1, Tiffanie R Johnson, Joseph D Tobias.   

Abstract

The authors present a 25-year-old woman who was admitted to the ICU for treatment of shock, respiratory failure, and acidosis related to congenital lactic acidosis from pyruvate dehydrogenase deficiency. To aid in ongoing management of the metabolic acidosis, the Paratrend blood gas monitoring sensor was inserted through a peripheral venous site to provide a continuous measurement of pH and partial pressure of carbon dioxide (Pco2). With the venous insertion of the Paratrend, a clinically useful correlation with arterial blood gas values was noted. Linear regression analysis of the pH values from the venous blood gas analyses and the Paratrend monitor revealed r2 = 0.71 with p = 0.001 and r2 = 0.78 with a p = 0.0003 for the Pco2 values. Our preliminary experience suggests that venous placement of the Paratrend monitor can be used to provide clinically useful, continuous measurement of pH and Pco2.

Entities:  

Mesh:

Year:  2002        PMID: 12086201     DOI: 10.1177/000992280204100508

Source DB:  PubMed          Journal:  Clin Pediatr (Phila)        ISSN: 0009-9228            Impact factor:   1.168


  2 in total

1.  Encapsulation of FITC to monitor extracellular pH: a step towards the development of red blood cells as circulating blood analyte biosensors.

Authors:  Sarah C Ritter; Mark A Milanick; Kenith E Meissner
Journal:  Biomed Opt Express       Date:  2011-06-22       Impact factor: 3.732

2.  Affection of the Respiratory Muscles in Combined Complex I and IV Deficiency.

Authors:  Josef Finsterer; Helmut Rauschka; Liane Segal; Gabor G Kovacs; Boris Rolinski
Journal:  Open Neurol J       Date:  2017-01-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.