| Literature DB >> 12083252 |
Wendy A Lepper1, Ann M Schultz, Michael S Curiale, Ronald L Johnson.
Abstract
The VIDAS Immuno-concentration Salmonella (ICS) plus selective plate method (Hektoen enteric, xylose lysine desoxycholate, bismuth sulfite) method for the detection of Salmonella was compared to the Bacteriological Analytical Manual (BAM)/AOAC culture method in a collaborative study. Thirty-two laboratories participated in the evaluation. Each laboratory tested one or more of the 6 test products: milk chocolate, nonfat dry milk, dried whole egg, soy flour, ground black pepper, and ground raw turkey. The 2 methods were in agreement for 1,297 of the 1,455 samples. Of the 158 samples not in agreement, 82 were VIDAS ICS plus selective plate-positive and BAM/AOAC-negative, and 76 were VIDAS ICS plus selective plate-negative and BAM/AOAC-positive.Entities:
Mesh:
Year: 2002 PMID: 12083252
Source DB: PubMed Journal: J AOAC Int ISSN: 1060-3271 Impact factor: 1.913