Two different fixations for treatment of Jones' fracture were tested in bone models and cadaveric specimens to determine the differences in the stability of the constructs. A bicortical 3.5-mm cannulated cortical screw and an intramedullary 4.0-mm partially threaded cancellous screw were tested using physiologic loads with an Instron 8500 servohydraulic tensiometer (Instron Corporation, Canton, MA). In bone models, the bicortical construct (n = 5, 87+/-23 N) showed superior fixation strength (p = .0009) when compared to the intramedullary screw fixation (n = 5, 25+/-13 N). Cadaveric testing showed similar statistical significance (p = .0124) with the bicortical construct (n = 5, 152+/-71 N) having greater load resistance than the intramedullary screw fixation (n = 4, 29+/-20 N). In bone models, the bicortical constructs (23+/-9 N/mm) showed over twice the elastic modulus than the intramedullary screw fixations (9+/-4 N/mm) with statistical significance (p = .0115). The elastic modulus in the cadaveric group showed a similar pattern between the bicortical (19+/-17 N/mm) and intramedullary (9+/-6 N/mm) screw constructs. Analysis of the bicortical screw failure patterns revealed that screw orientation had a critical impact on fixation stability. The more distal the exit site of the bicortical screw was from the fracture site, the greater the load needed to displace the fixation.Entities: