Literature DB >> 12067284

Artefactually-normal automated platelet counts due to malaria-infected RBC.

G Crabbe1, M Van Poucke, B Cantinieaux.   

Abstract

Protein aggregates, red cell or white cell fragments are known to interfere with platelet counts in automated blood analysers, both by aperture impedance and optical technologies. When a falsely high value is suspected, interference by pseudo-platelet particles can be confirmed by systematic examination of stained blood films. The method that best avoids these sources of interference is the reference, immunological platelet count. We describe a case of treated malaria with a false normal platelet count. The blood smear revealed small red cells, infected by trophozoites of Plasmodium falciparum, that interfered with the platelet count. The Cell Dyn 4000 shows different patterns of interference by infected red cells in its impedance and optical counts, and thrombocytopenia was suspected immediately. This was confirmed by a phase-contrast microscopic platelet count.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12067284     DOI: 10.1046/j.1365-2257.2002.00414.x

Source DB:  PubMed          Journal:  Clin Lab Haematol        ISSN: 0141-9854


  2 in total

1.  Artifactual changes in Sprague-Dawley rat hematologic parameters after storage of samples at 3 °C and 21 °C.

Authors:  Michelle C Cora; Debra King; Laura J Betz; Ralph Wilson; Greg S Travlos
Journal:  J Am Assoc Lab Anim Sci       Date:  2012       Impact factor: 1.232

Review 2.  Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution.

Authors:  Gene Gulati; Guldeep Uppal; Jerald Gong
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

  2 in total