| Literature DB >> 12064534 |
Joris Roels1, Herman Van Langenhove, Willy Verstraete.
Abstract
A gas chromatographic (GC) system to measure free phosphine in biogas and matrix bound phosphine in manure and sludge is presented. The system consists of a sample preconcentration trap filled with glass beads, connected with a capillary GC equipped with a thermionic specific detector. With a trap temperature as low as -155 degrees C, a sampling flow of 20 ml/min and a typical total sample volume of 100 ml, free phosphine concentrations in the low ng/m3 range and matrix bound phosphine in the low ng/kg dry matter range, can be accurately and reproducibly determined.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12064534 DOI: 10.1016/s0021-9673(02)00084-5
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759