Literature DB >> 12058426

[Usefulness of imaging method without using lead plate for sentinel lymph node scintigraphy].

Hiroyuki Tsushima1, Takashi Yamanaga, Yoshihiro Shimonishi, Hironobu Ochi.   

Abstract

PURPOSE: In scintigraphy of the sentinel lymph node, it is common to use a lead plate as a shield to reduce star artifact and scattered radiation in the portion surrounding the injected site of radioactive colloid. We have developed an imaging method without using a lead plate, and examined its usefulness in phantom and clinical imagings. METHODS,
RESULTS: Star artifact was eliminated using a medium energy collimator. Effects of scattered radiation from the injection site were reduced by setting energy window at higher level.
CONCLUSIONS: Our method without using a lead plate can be applied to dynamic data acquisition and imaging of a portion where it is difficult to place a lead plate. It also seems to be a useful imaging method in that it takes patients' feelings into consideration.

Entities:  

Mesh:

Year:  2002        PMID: 12058426

Source DB:  PubMed          Journal:  Kaku Igaku        ISSN: 0022-7854


  2 in total

1.  Efficacy of high-energy collimator for sentinel node lymphoscintigraphy of early breast cancer patients.

Authors:  Kamran Aryana; Mohaddeseh Gholizadeh; Mehdi Momennezhad; Maryam Naji; Mohsen Aliakbarian; Mohammad Naser Forghani; Ramin Sadeghi
Journal:  Radiol Oncol       Date:  2012-02-06       Impact factor: 2.991

2.  Lymphoscintigraphy and triangulated body marking for morbidity reduction during sentinel node biopsy in breast cancer.

Authors:  Borys R Krynyckyi; Michail K Shafir; Suk Chul Kim; Dong Wook Kim; Arlene Travis; Renee M Moadel; Chun K Kim
Journal:  Int Semin Surg Oncol       Date:  2005-11-08