| Literature DB >> 12051301 |
Abstract
The findings of clinical pilot study (n = 9 subjects) using a new laser Doppler sensor for assessing blood flux in plantar skin tissue are described. Cutaneous blood perfusion was recorded under the first metatarsal head (right foot) in standing and walking. The sensor was located in a measurement shoe custom made for each test subject. The test group comprised diabetic patients (type II) with vascular (n = 3) or neuropathic (n = 3) complications and three controls. All subjects were Caucasian males and in the age range considered particularly at risk of diabetic foot problems (mean 61 years, 51-72 years). Following static loading for 2, 3 and 4 min the blood.flux response increased rapidly in the control (mean = 10 s) and neuropathic (mean = 18 s) groups to a well-defined, peak blood flow. For the vascular group. the blood flux response was typically a slower rise (mean = 30 s) to a poorly defined peak blood flow value. Due to movement artifact a reliable signal could only be obtained for the swing phase of gait during which blood flux was observed to increase linearly. This was interpreted as reperfusion of the tissue following unloading. The rates of reperfusion expressed in arbitrary units (of blood flux) per millisecond (au ms(-1)) were 6.1-7.9 au ms(-1) for the control, 4-6.2 au ms- for the vascular and 2.3-4.5 au ms(-1) for the neuropathic groups. The feasibility of assessing the microcirculation of the plantar skin under conditions of static and dynamic loading, with the foot in-shoe, has been demonstrated for the first time. The results suggest that abnormal responses may be obtained from asymptomatic feet of diabetic patients with vascular and/or neuropathic complications. This method of assessment could be of use in predicting the occurrence of ulceration in the diabetic foot.Entities:
Mesh:
Year: 2002 PMID: 12051301 DOI: 10.1088/0967-3334/23/2/305
Source DB: PubMed Journal: Physiol Meas ISSN: 0967-3334 Impact factor: 2.833