Ana Paula S Paim1, Cristina M N V Almeida, Boaventura F Reis, Rui A S Lapa, Elias A G Zagatto, José L F Costa Lima.
Abstract
A flow procedure for the determination of ascorbic acid in pharmaceutical formulations exploiting potentiometric titration is described. The method is based on the reduction of IO3- by ascorbic acid and the detection was carried out employing a flow-through ion selective electrode for iodide. The flow network controlled by a microcomputer was designed to implement multicommutation for ease of operation and robustness. The titration system allowed the determination of ascorbic acid in pharmaceutical formulations with concentrations ranging from 7.5 to 15.0 mmol l(-1). No significant differences at the 95% confidence level were observed in comparison with results obtained by a manual procedure. Merit figures of results such as a relative standard deviation of 1.0% (n=6) and a reagent consumption of 21.4 mg IO3- per determination were obtained.Entities: