Severe hemolysis due to cloth wear 23 years after aortic valve replacement on a Starr-Edwards ball valve model 2320.

Mikiko Murakami1, Hiroyuki Tanaka, Masazumi Watanabe, Masato Shimizu, Makoto Sunamori.   

Abstract

Despite iron substitution therapy, a patient developed severe hemolytic anemia 23 yr after insertion of a cloth-covered Starr-Edwards model 2320 aortic valve prosthesis. The prosthesis showed no sign of significant dysfunction. Upon removal, it showed extensive cloth wear on the inner surface of all three struts; one strut was completely denuded of its cloth covering. Hemolysis immediately resolved after replacement with a St Jude aortic prosthesis.

Entities:  

Mesh:

Year:  2002        PMID: 12044439     DOI: 10.1016/s0967-2109(01)00130-2

Source DB:  PubMed          Journal:  Cardiovasc Surg        ISSN: 0967-2109