Literature DB >> 12036652

Microphacoemulsification with WhiteStar. A wound-temperature study.

William Soscia1, James G Howard, Randall J Olson.   

Abstract

PURPOSE: To verify the temperature-induction effect of bimanual ultrasound-based phacoemulsification through 2 stab incisions with a bare aspiration needle and new proprietary technology.
SETTING: In vitro laboratory.
METHODS: The Sovereign phacoemulsification unit and the WhiteStar system (Allergan) were used in conjunction with the Olson irrigating chopper (ASICO) and a 19-gauge bare phaco needle in 2 human cadaver eye-bank eyes at 100% power unoccluded, 100% power with aspiration completely occluded, and 100% power with aspiration and flow completely occluded. Temperature data were monitored throughout the experiment. Any sign of a wound burn, defined as temperature exceeding 45 degrees C, whitening around the wound, or wound contraction, was the end point.
RESULTS: At 100% continuous phacoemulsification power, the temperature did not rise above 27.3 degrees C. When the aspiration line was totally occluded at 100% continuous power, the temperature did not rise above 32.5 degrees C. When flow into the eye and aspiration were completely occluded at 100% power, a wound change and temperature of 45 degrees C occurred in 45 seconds in the first eye and in 29 seconds in the second eye.
CONCLUSIONS: Using the Sovereign phacoemulsification unit with the WhiteStar system during 2-stab-incision bimanual microphacoemulsification with a bare 19-gauge aspiration needle in human cadaver eyes, a wound burn could not be produced at the highest energy settings unless all flow into the eye and all aspiration were occluded. These settings were well beyond clinically applicable conditions and provide evidence that microphacoemulsification can be performed safely with the WhiteStar system.

Entities:  

Mesh:

Year:  2002        PMID: 12036652     DOI: 10.1016/s0886-3350(02)01276-2

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Bimanual microincisional cataract surgery technique and clinical outcome.

Authors:  Abdulrahman Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-08-05

2.  [Technologies, techniques and tactics for micro-incision cataract surgery].

Authors:  E Fabian; M Maier
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

3.  Heat profiling of phacoemulsification tip using a thermal scanning camera.

Authors:  Wei Kiong Ngo; Louis W Lim; Colin S H Tan; Wee Jin Heng
Journal:  Int Ophthalmol       Date:  2013-03-20       Impact factor: 2.031

4.  Effect of Corneal Incision Enlargement on Surgically Induced Astigmatism in Biaxial Microincision Cataract Surgery.

Authors:  Mehmet Tetikoğlu; Celal Yeter; Fırat Helvacıoğlu; Serdar Aktaş; Hacı Murat Sağdık; Fatih Özcura
Journal:  Turk J Ophthalmol       Date:  2016-06-06

5.  Insertion of a foldable hydrophobic IOL through the trabeculectomy fistula in cases with microincision cataract surgery combined with trabeculectomy.

Authors:  Tanuj Dada; Rajamani Muralidhar; Harinder S Sethi
Journal:  BMC Ophthalmol       Date:  2006-03-19       Impact factor: 2.209

6.  Comparison of Floaters after Cataract Surgery with Different Viscoelastics.

Authors:  Jinsoo Kim; Hak Jun Lee; In Won Park; Soon Il Kwon
Journal:  Int J Med Sci       Date:  2018-01-01       Impact factor: 3.738

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.