Endograft fabric disintegration simulating a type II endoleak.

Jean-Pierre Becquemin1, Bertrand Poussier, Eric Allaire, Hischam Kobeiter, Pascal Desgranges.   

Abstract

PURPOSE: To report a case of endograft disintegration misinterpreted as a type II endoleak. CASE REPORT: An 86-year-old man underwent successful Vanguard stent-graft implantation for a 49-mm abdominal aortic aneurysm. At 1 month, a distal type I endoleak was repaired, but another endoleak thought to arise from the lumbar or inferior mesenteric arteries remained. Multiple interventions to embolize the feeding arteries were unsuccessful, and the endoleak persisted. At 1 year, the leak was still present, and the aneurysm had increased to 69 mm. After failing to demonstrate the source of the leak radiographically, surgery was performed. At operation, the endoleak was traced to small holes in the graft fabric. The endograft was removed without difficulty, and the flow was re-established with an aortobi-iliac woven graft.
CONCLUSIONS: Chronically implanted stent-grafts can show signs of failure that are confusing or misleading. Correct diagnosis of endoleaks may be difficult, but every effort must be made to identify their source. It is vital that no enlarging aneurysm be left untreated.

Entities:  

Mesh:

Year:  2002        PMID: 12010101     DOI: 10.1177/152660280200900211

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  2 in total

1.  Late Type 3b Endoleak Mimicking Type 2 Endoleak after Endovascular Aortic Aneurysm Repair.

Authors:  Minji Cho; Chanjoong Choi; Sungsin Cho; Song-Yi Kim; Sang-Il Min; Sanghyun Ahn; Jongwon Ha; Seung-Kee Min
Journal:  Vasc Specialist Int       Date:  2017-06-30

2.  Type III B endoleak leading to aortic rupture after endovascular repair: analysis of errors in follow up and treatment.

Authors:  Marco Leopardi; Alessia Salerno; Pietro Scarpelli; Marco Ventura
Journal:  CVIR Endovasc       Date:  2018-11-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.