| Literature DB >> 12008134 |
Florentin Tache1, Alexandru Farca, Andrei Medvedovici, Victor David.
Abstract
Both derivatization of free captopril in human plasma samples using monobromobimane as fluorescent label and the corresponding HPLC-fluorescence detection (FLD) method were validated. Calibration curve for the fluorescent captopril derivative in plasma samples is linear in the ppb-ppm range with a detection limit of 4 ppb and an identification limit of 10 ppb (P%: 90; nu > or = 5). These methods were successfully applied on bioequivalence studies carried out on some marketed pharmaceutical formulations.Entities:
Mesh:
Substances:
Year: 2002 PMID: 12008134 DOI: 10.1016/s0731-7085(01)00687-2
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935