Error analysis of tissue resistivity measurement.

Jang-Zern Tsai1, James A Will, Scott Hubbard-Van Stelle, Hong Cao, Supan Tungjitkusolmun, Young Bin Choy, Dieter Haemmerich, Vicken R Vorperian, John G Webster.   

Abstract

We identified the error sources in a system for measuring tissue resistivity at eight frequencies from 1 Hz to 1 MHz using the four-terminal method. We expressed the measured resistivity with an analytical formula containing all error terms. We conducted practical error measurements with in-vivo and bench-top experiments. We averaged errors at all frequencies for all measurements. The standard deviations of error of the quantization error of the 8-bit digital oscilloscope with voltage averaging, the nonideality of the circuit, the in-vivo motion artifact and electrical interference combined to yield an error of +/- 1.19%. The dimension error in measuring the syringe tube for measuring the reference saline resistivity added +/- 1.32% error. The estimation of the working probe constant by interpolating a set of probe constants measured in reference saline solutions added +/- 0.48% error. The difference in the current magnitudes used during the probe calibration and that during the tissue resistivity measurement caused +/- 0.14% error. Variation of the electrode spacing, alignment, and electrode surface property due to the insertion of electrodes into the tissue caused +/- 0.61% error. We combined the above errors to yield an overall standard deviation error of the measured tissue resistivity of +/- 1.96%.

Entities:  

Mesh:

Year:  2002        PMID: 12002180     DOI: 10.1109/10.995687

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  5 in total

1.  A new approach for resolution of complex tissue impedance spectra in hearts.

Authors:  Andrew E Pollard; Roger C Barr
Journal:  IEEE Trans Biomed Eng       Date:  2013-04-18       Impact factor: 4.538

2.  Electrical conductivity measurement of excised human metastatic liver tumours before and after thermal ablation.

Authors:  Dieter Haemmerich; David J Schutt; Andrew W Wright; John G Webster; David M Mahvi
Journal:  Physiol Meas       Date:  2009-04-06       Impact factor: 2.833

3.  Electrical impedance along connective tissue planes associated with acupuncture meridians.

Authors:  Andrew C Ahn; Junru Wu; Gary J Badger; Richard Hammerschlag; Helene M Langevin
Journal:  BMC Complement Altern Med       Date:  2005-05-09       Impact factor: 3.659

4.  Myocardial electrotonic response to submaximal exercise in dogs with healed myocardial infarctions: evidence for β-adrenoceptor mediated enhanced coupling during exercise testing.

Authors:  Carlos L Del Rio; Bradley D Clymer; George E Billman
Journal:  Front Physiol       Date:  2015-02-05       Impact factor: 4.566

5.  Probe Contact Force Monitoring during Conductivity Measurements of the Left Atrial Appendage to Support the Design of Novel Diagnostic and Therapeutic Procedures.

Authors:  Hamza Benchakroun; Niko Ištuk; Eoghan Dunne; Muhammad Adnan Elahi; Tony O'Halloran; Martin O'Halloran; Declan O'Loughlin
Journal:  Sensors (Basel)       Date:  2022-09-21       Impact factor: 3.847

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.