Literature DB >> 11999197

The Aachen-keratoprosthesis--a flexible KPro that permits intraocular pressure measurement.

A Krug1, S Kompa, N F Schrage.   

Abstract

PURPOSE: Postoperative glaucoma is one of the major complications observed in patients with keratoprostheses (KPro). The optical components of the majority of KPros are rigid and inflexible, which prevents indirect tonometry by common methods. This study confirms that the Aachen-KPro allows measurements of intraocular pressure (IOP) due to its flexible optical part with a modified Schiotz tonometer.
METHODS: The Aachen-KPro was placed in a special chamber where pressure can be generated and monitored by a transducer. Measurements were taken by common ophthalmologic methods. The results were compared to the initial pressure values in the chamber.
RESULTS: With the Goldmann tonometer, the Tono-Pen and the topography system (Technomed), no significant pressure changes could be observed. The results of the Schiotz tonometer were promising. It was slightly modified and standard curves with different weights could be obtained.
CONCLUSIONS: With a modified Schiotz tonometer, it is possible to detect elevated IOPs. Modifying this common instrument is inexpensive and can be easily performed. This advantage of the Aachen-KPro permits early management of postoperative high tension glaucoma.

Entities:  

Mesh:

Year:  2002        PMID: 11999197     DOI: 10.1177/039139880202500311

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  3 in total

1.  Five year follow up of biocolonisable microporous fluorocarbon haptic (BIOKOP) keratoprosthesis implantation in patients with high risk of corneal graft failure.

Authors:  J L Alió; M E Mulet; H Haroun; J Merayo; J M Ruiz Moreno
Journal:  Br J Ophthalmol       Date:  2004-12       Impact factor: 4.638

2.  [In vitro and in vivo investigations of a modified textile keratoprosthesis. ACTO TexKPRO].

Authors:  N Schrage; A K Cordes; J Storsberg; S Sel; H Röthgen; N Nass; E Berndt; S Jockenhövel; A Meßner; C Panfil
Journal:  Ophthalmologe       Date:  2014-03       Impact factor: 1.059

Review 3.  Challenges of Glaucoma Management in Patients with Type I Boston Keratoprosthesis.

Authors:  Sara M AlHilali; Samar A Al-Swailem
Journal:  Clin Ophthalmol       Date:  2022-02-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.