EUS guidance in gastric pacemaker implantation.

Gottumukkala S Raju1, Jameson Forster, Irene Sarosiek, Stanton J Rosenthal, Zhiyue Lin, Richard McCallum.   

Abstract

BACKGROUND: EUS provides excellent imaging of the gastric wall. The utility of EUS imaging in guiding the placement of a gastric pacemaker was investigated.
METHODS: Fourteen patients underwent gastric pacemaker implantation for refractory gastroparesis at laparotomy. Placement of the lead into the muscle layer of the antrum of the stomach was imaged by intraoperative surface ultrasonography in the first 8 patients and by EUS in the subsequent 6 patients.
RESULTS: Surface US examination of the lead placement revealed reverberation artifacts. The images were uniformly unsatisfactory and the position of the lead in the gastric wall could not be visualized in any patient. In contrast, the lead was clearly and easily identified by EUS as a bright linear echo in the gastric wall. This was observed uniformly in all of the patients evaluated by EUS. Compared with surface US, EUS provided better images of the gastric lead placement as well as less abdominal distension and thus easier closure of the incision.
CONCLUSIONS: EUS is useful in confirming the accurate placement of pacemaker leads within the muscular coat of the stomach.

Entities:  

Mesh:

Year:  2002        PMID: 11979260     DOI: 10.1067/mge.2002.123275

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  1 in total

1.  Gastric electrical stimulation for gastroparesis: a goal greatly pursued, but not yet attained.

Authors:  Mauro Bortolotti
Journal:  World J Gastroenterol       Date:  2011-01-21       Impact factor: 5.742

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.