Literature DB >> 11967592

Cardiac output measurement by pulse dye densitometry: a comparison with the Fick's principle and thermodilution method.

Frank Bremer1, Albert Schiele, Klaus Tschaikowsky.   

Abstract

OBJECTIVE: To evaluate the agreement between cardiac output (CO) measurements obtained by a new dye dilution technique using pulse dye densitometry (PDD) and thermodilution (TD) and the direct Fick method (F). DESIGN AND
SETTING: Prospective clinical study in a university hospital, cardiac surgery intensive care unit. PATIENTS: Fifty-eight cardiac surgery patients after admission to the intensive care unit (six were excluded due to a low pulse signal quality using the PDD method). MEASUREMENTS AND
RESULTS: Mean CO was 5.3+/-1.8 l/min for PDD, 5.7+/-1.68 l/min for TD, and 6.16+/-1.66 l/min for F. There was a good correlation between PDD and TD ( r(2)=0.93) and between PDD and F ( r(2)=0.77). Bias and precision between PDD and TD were -0.39+/-0.5 l/min and -0.69+/-0.85 l/min between PDD and F. In general, PDD determined lower CO values than TD and F. Especially in patients with CO below 5 l/min PDD underestimated CO in comparison to TD and F (bias and precision: -0.51+/-0.40 l/min and -0.83+/-1.0 l/min).
CONCLUSION: Comparison between PDD and TD showed good agreement for the normal to high CO range. However, agreement was poor in patients with low CO. In the latter patient group PDD showed relevant underestimation of CO compared to TD and F. Due to these limitations PDD cannot entirely replace the pulmonary artery catheter for CO determination.

Entities:  

Mesh:

Year:  2002        PMID: 11967592     DOI: 10.1007/s00134-002-1252-3

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  8 in total

1.  Neurotoxic effects of indocyanine green -cerebellar granule cell culture viability study.

Authors:  Beata Toczylowska; Elzbieta Zieminska; Grazyna Goch; Daniel Milej; Anna Gerega; Adam Liebert
Journal:  Biomed Opt Express       Date:  2014-02-19       Impact factor: 3.732

2.  Cardiac output measurement by pulse dye densitometry: comparison with pulmonary artery thermodilution in post-cardiac surgery patients.

Authors:  Martijn Kroon; A B Johan Groeneveld; Yvo M Smulders
Journal:  J Clin Monit Comput       Date:  2006-01-25       Impact factor: 2.502

3.  Comparison of oxygen consumption calculated by Fick's principle (using a central venous catheter) and measured by indirect calorimetry.

Authors:  Chiaki Inadomi; Yoshiaki Terao; Kazunori Yamashita; Makoto Fukusaki; Masafumi Takada; Koji Sumikawa
Journal:  J Anesth       Date:  2008-05-25       Impact factor: 2.078

4.  In vivo multispectral photoacoustic and photothermal flow cytometry with multicolor dyes: a potential for real-time assessment of circulation, dye-cell interaction, and blood volume.

Authors:  Mikhail A Proskurnin; Tatyana V Zhidkova; Dmitry S Volkov; Mustafa Sarimollaoglu; Ekaterina I Galanzha; Donald Mock; Dmitry A Nedosekin; Vladimir P Zharov
Journal:  Cytometry A       Date:  2011-09-08       Impact factor: 4.355

5.  Stability of cirrhotic systemic hemodynamics ensures sufficient splanchnic blood flow after living-donor liver transplantation in adult recipients with liver cirrhosis.

Authors:  Tomohide Hori; Shintaro Yagi; Taku Iida; Kentaro Taniguchi; Kentaro Yamagiwa; Chiduru Yamamoto; Takashi Hasegawa; Koichiro Yamakado; Takuma Kato; Kanako Saito; Linan Wang; Mie Torii; Yukinobu Hori; Kan Takeda; Kazuo Maruyama; Shinji Uemoto
Journal:  World J Gastroenterol       Date:  2007-11-28       Impact factor: 5.742

6.  [Noninvasive determination of cardiac output in ventilated patients . Clinical evaluation of a simplified quick method].

Authors:  J Reutershan; T Kapp; K Unertl; R Fretschner
Journal:  Anaesthesist       Date:  2003-09       Impact factor: 1.041

7.  Reliability of cardiac output calculation by the fick principle and central venous oxygen saturation in emergency conditions.

Authors:  Avi A Weinbroum; Philippe Biderman; Dror Soffer; Joseph M Klausner; Oded Szold
Journal:  J Clin Monit Comput       Date:  2008-10-23       Impact factor: 2.502

8.  Basic and clinical assessment of initial distribution volume of glucose in hemodynamically stable pediatric intensive care patients.

Authors:  Hironori Ishihara; Eiji Hashiba; Hirobumi Okawa; Junichi Saito; Toshinori Kasai; Toshihito Tsubo
Journal:  J Intensive Care       Date:  2014-11-12
  8 in total